EXCELL PRECISION CO., LTD. Table of Contents 1. Appearance and Part Descriptions ..................5 2. Power Supply ......................... 6 3. Connector/Cable Definitions ....................6 4. Before Use ..........................7 5. Use TM with GoW/QoW ......................8 6. Use TM with APP ........................8 6-1.
EXCELL PRECISION CO., LTD. 2. Power Supply Power selection and consumption Lithium battery: DC 3.7V / 1500mAh Adapter: Input: 100~240VAC (50~60Hz) Output: 5VDC 2A Connected to one 350Ω load cell, 25°C Status Power consumption Battery life Working...
EXCELL PRECISION CO., LTD. 5. Use TM with GoW/QoW Pairing/Calibration Hold [Re-pairing/Clear] key until Link/Pair LED flashing. Press [Net] and [ON/OFF] keys together to enter self test mode. Press [Zero] or [Unit] key to select”06RePr” and then press [Tare/PT] key to pair. Link/Pair LED always on means pairing is successful, otherwise repeat.
EXCELL PRECISION CO., LTD. 3 . Connect the Device Run the APP and select [WEIGHT] menu Click [Platform no] to enter device list. Click the device to be connected. Once connected, weight value and device name display on app.
Page 12
EXCELL PRECISION CO., LTD. 3 . Connect the Device Run the APP and select [WEIGHT] menu Click [Platform no] to enter device list. Click the device to be connected. Once connected, weight value and device name display on app.
Page 13
EXCELL PRECISION CO., LTD. (If Linearity Segment set as 0,only do ~) No load on platform, click [Confirm]. Place 1/2 of Max. Capacity of weight on platform and click [Confirm]. Place Max. Capacity of weight on platform and click [Confirm].
EXCELL PRECISION CO., LTD. 7. Key Description Stable: weight shows in “white”, Unstable: weight shows in “black” Zero key: When the weight is stable and is not zero, click this key to zero the scale.
EXCELL PRECISION CO., LTD. AD RAW DATA key: Click [AD RAW DATA] to shift between weight and internal value. Click [AD RAW DATA] for several secs to set this key as Temporary/Permanent, display selection for AD and display selection for weight. Click [CONFIRM] to finish setting.
EXCELL PRECISION CO., LTD. 8-2. Change the Device Name After connecting TM and APP, run the APP and select [WEIGHT] menu Press and hold the device name such as " ExBLE-P " until the dialouge shows up. E.g “ExBLE-P”...
EXCELL PRECISION CO., LTD. 8-4. Set/Change Password (for Password Calibration) After setting TM to Calibration Mode and connecting the APP to TM successfully, press the “Password” button, and the Password dialouge will appear. Enter a new password in the Password dialouge and click [Confirm].
EXCELL PRECISION CO., LTD. 8-6. Password Calibration After you have enabled Password Calibration in Function Setting dialogue (8-5), set a password in the Password dialogue (8-4), and connected the APP to TM successfully, press the “Password Calibration” button, and the Password Input dialouge will appear.
EXCELL PRECISION CO., LTD. 9. Bluetooth UUID Description and Data Formats 9-1. UUID Description UUID E812 : Used for setting Bluetooth and calibrating scale. The Commandss used are as follows: (1) Calibration Commands 0x99 read/write; (2) Password Commands 0x9C read/write UUID E813 : Scale weight or internal value output, scale keypad input (please refer to 11.
EXCELL PRECISION CO., LTD. 9-2-4. Data Type B Data Type B is the ordinary base-16 data type. All 8 bits of one byte are data. Example : Data =5000 converted to 4-bytes data format would be Type B(Hex) = 00 00 13 88 Data =255 converted to 2-bytes data format would be Type B(Hex) = 00 FF 9-2-5.
Page 21
EXCELL PRECISION CO., LTD. C1: Scale Calibration Mode (enable/disable) Description Scale Calibration Mode enabled Normal Mode P1: Scale Power Status (ON/OFF) Description Scale power ON Scale power OFF R1: “Ready to Pair” State (R1 bit is functional only when Z1 bit is 0) Description “Ready to Pair”...
EXCELL PRECISION CO., LTD. 11. Bluetooth Weight Transmission Formats (for UUID E813 Handle) The UUID E813 Notify Bluetooth transmission format has 3 AD internal value modes and 3 weight modes. For enabling these modes please refer to 12-1-1. Temporarily Switch E813 Notify Output Formats and 12-1-22.
Page 23
EXCELL PRECISION CO., LTD. D1 ~ D3: Power Status Description Externally powered model, power source voltage normal Externally powered model, power source voltage too low Lithium-ion battery-powered model, battery in use Lithium-ion battery-powered model, charging battery Lithium-ion battery-powered model, battery fully charged...
EXCELL PRECISION CO., LTD. R1~R24: AD internal value R1~R24 Description R24~R1 AD internal value ,R24 is MSB, R1 is LSB 11-2 Weight Mode Settings When the following commands are written to UUID E812, UUID E813 Notify’s Weight Mode can be changed;...
Newline 0D 0A (2 Bytes Hex data) 11-3-2 Weight Mode 1 Input Commands (Weight Mode 1 is for special use by some Excell models. Normal users are recommended to use Weight Mode 2 or 3.) The following commands perform “Characteristics Value Write (Write Request)” for UUID E813 handle:...
EXCELL PRECISION CO., LTD. 11-4 Weight Mode 2 11-4-1 Weight Mode 2 Output Formats According to 11-2 Weight Mode Settings along with Weight Mode 2 setting, when UUID E813 Notify is enabled, the following table’s Weight Mode 2 values can be read:...
Page 28
EXCELL PRECISION CO., LTD. Grain Penny Weight Carat Hong Kong Tael Troy Ounce Reserved Z1: Zero Point Indicator Description Zero point indicator ON Zero point indicator OFF O1: Scale Off Flag Description Scale OFF Scale ON N1: Net Weight Indicator...
Page 29
EXCELL PRECISION CO., LTD. D1 ~ D3: Power Status Description Externally powered model, power source voltage normal Externally powered model, power source voltage too low Lithium-ion battery-powered model, battery in use Lithium-ion battery-powered model, charging battery Lithium-ion battery-powered model, battery fully charged...
EXCELL PRECISION CO., LTD. A1~A6 : 6-byte ASCII Code, pre-tare value without decimal point Example : If you receive E813 Notify’s pre-tare value return format as 9B 09 01 03 08 20 20 31 30 30 30 , that means the pre-tare value is 1.000kg.
EXCELL PRECISION CO., LTD. P1~P6: Decimal Point Position P6 P5 P4 P3 P2 P1 Description No decimal point Displays 0. , decimal point after the number (no decimal places) Displays 0.0 , one decimal place Displays 0.00 , two decimal places Displays 0.000 , three decimal places...
EXCELL PRECISION CO., LTD. Data String Description Unit (3 Bytes), Unit Formats : <SP> denotes SPACE character, 20 (1 byte Hex data) No Unit <SP> SP> SP> Milligram <SP> m Gram <SP> <SP> g Unit Byte – 18 Byte Kilogram <SP>...
EXCELL PRECISION CO., LTD. 12-1-3. Write Gravity Value at Calibration This 42(Hex) sub-command set can be written only when Calibration Mode is enabled. Write UUID E812 handle (Hex) with the following command(s): Calibration Length Sub-Command 4-byte Type A Data Command ‘Write’...
EXCELL PRECISION CO., LTD. 12-1-5. Write User Location G Value This 43(Hex) sub-command set can be written only when Calibration Mode is enabled. Write UUID E812 handle (Hex) with the following command(s): Calibration Length Sub-Command 4-byte Type A Data ‘Write’ Description...
EXCELL PRECISION CO., LTD. 12-1-7. Write Maximum Displayed Weighing Capacity This 44(Hex) sub-command set can be written only when Calibration Mode is enabled. Write UUID E812 handle (Hex) with the following command(s): Calibration Length Sub-Command 4-byte Type A Data ‘Write’ Description...
EXCELL PRECISION CO., LTD. 12-1-9. Write Division This 45(Hex) sub-command set can be written only when Calibration Mode is enabled. Write UUID E812 handle (Hex) with the following command(s): Calibration Length Sub-Command 1-byte Type A Data ‘Write’ Description Command (Hex)
EXCELL PRECISION CO., LTD. Execution conditions for these Calibration Mode Normal Mode commands Executable Not executable 12-1-12. Read Decimal Point Position for Display Write UUID E812 handle (Hex) with the following command(s): Calibration Command (Hex) Length (Hex) Sub-Command (Hex) After the above ‘Read’ command set is successfully written to the UUID E812 handle, there will be...
EXCELL PRECISION CO., LTD. Read UUID E812 handle, which will return 99 02 7F ..(please refer to 12-1-39 Return Read & Write Status). Example : For 10kg weight calibration, empty-load AD value reading after filtering is 2000; 10kg-load AD value after filtering is 202000;...
EXCELL PRECISION CO., LTD. 12-1-18. Write Zero Range This 4A(Hex) sub-command set can be written only when Calibration Mode is enabled. Write UUID E812 handle (Hex) with the following command(s): Calibration Length Sub-Command 1-byte Type A Data ‘Write’ Description Command...
EXCELL PRECISION CO., LTD. 12-1-20. Write Weight Unit This 4B(Hex) sub-command set can be written only when Calibration Mode is enabled. Write UUID E812 handle (Hex) with the following command(s): Calibration Length Sub-Command 1-byte Type A Data ‘Write’ Description Command...
EXCELL PRECISION CO., LTD. 12-1-22. Write Power-on E813 Notify Mode This 4C(Hex) sub-command set can be written only when Calibration Mode is enabled. Write UUID E812 handle (Hex) with the following command(s): Calibration Length Sub-Command 1-byte ‘Write’ Description Command (Hex)
EXCELL PRECISION CO., LTD. 12-1-26. Write Zero Tracking Range This 4E(Hex) sub-command set can be written only when Calibration Mode is enabled. Write UUID E812 handle (Hex) with the following command(s): Calibration Length Sub-Command 1-byte ‘Write’ Description Command (Hex) (Hex)
EXCELL PRECISION CO., LTD. 12-1-28. Write Scale Function Settings This command is supported only in ADP18 or its later versions, plus AFP15 or its later versions. This 4F(Hex) sub-command set can be written only when Calibration Mode is enabled. Write UUID E812 handle (Hex) with the following command(s):...
EXCELL PRECISION CO., LTD. 12-1-29. Read Scale Function Settings This command is supported only in ADP18 or its later versions, plus AFP15 or its later versions. Write UUID E812 handle (Hex) with the following command(s): Calibration Command (Hex) Length (Hex) Sub-Command (Hex) After the above ‘Read’...
EXCELL PRECISION CO., LTD. Read UUID E812 handle, which will return 99 02 7F ..(please refer to 12-1-39 Return Read & Write Status). Execution conditions for these Calibration Mode Normal Mode commands Executable Executable 12-1-31. Read Bluetooth Mode This command is supported only in ADP18 or its later versions, plus AFP15 or its later versions.
EXCELL PRECISION CO., LTD. After the above ‘Read’ command set is successfully written to the UUID E812 handle, there will be return values (Hex data) after reading the UUID E812 handle, such as the following: 99 03 11 TT TT (for descriptions regarding the TT return value(s), please refer to 12-1-32. Write Scale’s Power-off Delay (Minutes) after Disconnection).
EXCELL PRECISION CO., LTD. 12-1-36. Write Stabilization Speed Settings This command is supported only in ADP18 or its later versions, plus AFP15 or its later versions. Write UUID E812 handle (Hex) with the following command(s): Calibration Length Sub-Command Type A Data ‘Write’...
EXCELL PRECISION CO., LTD. Example: The return values are: 99 06 14 41 44 50 31 38 Then the firmware version is : 41 44 50 31 38 , i.e. ADP18 If the ‘Write’ operation has failed, reading UUID E812 handle will instead return 99 02 7F ..(please refer to 12-1-39 Return Read &...
EXCELL PRECISION CO., LTD. 12-2 Password Commands 9C(Hex) 12-2-1 Set <LINK PASSWORD> For this setting, you can set Connection Password <LINK PASSWORD> only when the H1 bit (Connection Password Decryption Required for Every Connection), as described in the section 12-1-31. Read Bluetooth Mode, is set to 1.
EXCELL PRECISION CO., LTD. 12-2-2 Input <LINK PASSWORD> to Complete Connection Decryption This decryption can be performed only when the H1 bit Connection Password Decryption Required for Every Connection), as described in the section 12-1-31. Read Bluetooth Mode , is set to 1.
EXCELL PRECISION CO., LTD. “Temporary Code” AES128 ECB Encryption Input (16 bytes) Key (16 bytes) Output (16 bytes) “Temporary Code” <LINK PASSWORD> NN NN NN NN NN NN NN NN NN NN NN NN NN NN NN NN Decrypt and get “Temporary Code”, please refer to the table below: “Temporary Code”...
EXCELL PRECISION CO., LTD. Use the <CAL PASSWORD> and Default Key to perform AES128 ECB Encryption to produce the 16-byte encrypted output XX XX XX XX XX XX XX XX XX XX XX XX XX XX XX XX Default Key is a fixed set of 16-byte Hex data = 0D 99 74 9F 3F 48 D1 9B 22 91 A2 77 24 44 74 9B <CAL PASSWORD>, if the byte count is less than 16 bytes, then prefix instance(s) of 00(Hex) to fill up 16...
Page 63
EXCELL PRECISION CO., LTD. (3). Read “Temporary Code”: For UUID E812 handle write the following Commands : Read “Temporary Code” Command (Hex) Length (Hex) Sub-Command (Hex) (4). Read UUID E812 handle and request decryption of “Temporary Code” Read UUID E812 handle, then you can get the following return value: “Temporary Code”...
EXCELL PRECISION CO., LTD. 13. Usage Descriptions TM box may be referred to as “this device” in this chapter. 13-1 Communication Protocol Communication Protocol is Bluetooth 4.0 Bluetooth Low Energy, BLE Protocol. Activate BLE scan on your mobile device or PC.
EXCELL PRECISION CO., LTD. 13-4 Connection Password The function of a connection password is to share this device with other particular devices without allowing unintended connections from everyone. If you want to require a password for connection, you can enable Connection Password Mode – for more details please refer to 12-1-30.
EXCELL PRECISION CO., LTD. 13-7 Calibration Password Calibration Password ’s function is that this device can decrypt Calibration Password via connection to enter Calibration Mode, but Calibration Mode will be closed after disconnection. To enable Calibration Password Mode , please refer to 12-1-28. Write Scale Function Settings and ensure that the W1 bit (Enable Calibration Mode via Input Password) is set to 1.
EXCELL PRECISION CO., LTD. 13-10 How to Calibrate Weight For the following procedure, be cautious about the conversion of Type A Data to be written (please refer to 9-2-3. Data Type A regarding Type A format(s)). ❶ First set this device to Calibration Mode.