EXCELL PRECISION CO., LTD.
(If Linearity Segment set as 0,only do ~)
No load on platform, click [Confirm].
Place 1/2 of Max. Capacity of weight on platform and click [Confirm].
Place Max. Capacity of weight on platform and click [Confirm].
Start weight calibration. Remove the weight on platform and click [Confirm].
Place Max. Capacity of weight on platform and click [Confirm].
Remove the weight on platform and click [Confirm] to finish calibration.
7. To end calibration mode, press [Calibration] key for 3 secs until calibration LED off.