EXCELL PRECISION CO., LTD.
12-1-9. Write Division
This 45(Hex) sub-command set can be written only when Calibration Mode is enabled.
Write UUID E812 handle (Hex) with the following command(s):
Calibration
Length
Sub-Command
Command
(Hex)
(Hex)
(Hex)
99
02
45
Read UUID E812 handle, which will return 99 02 7F .... (please refer to 12-1-39 Return Read & Write
Status)
Execution conditions for these
commands
12-1-10. Read Division
Write UUID E812 handle (Hex) with the following command(s):
Calibration Command (Hex)
99
After the above 'Read' command set is successfully written to the UUID E812 handle, there will be
return values (Hex data) after reading the UUID E812 handle, such as the following:
99 02 05 GG
GG: 1-byte Division; please refer to 12-1-9 Write Division
If the 'Write' operation has failed, reading UUID E812 handle will instead return 99 02 7F .... (please
refer to 12-1-39 Return Read & Write Status).
Execution conditions for these
commands
12-1-11. Write Decimal Point Position for Display
This 46(Hex) sub-command set can be written only when Calibration Mode is enabled.
Write UUID E812 handle (Hex) with the following command(s):
Calibration
Length
Sub-Command
Command
(Hex)
(Hex)
(Hex)
99
02
46
Read UUID E812 handle, which will return 99 02 7F .... (please refer to 12-1-39 Return Read & Write
Status).
TM
1-byte Type A Data
'Write' Description
(Hex)
GG
Division (available settings: 1,2,5)
Calibration Mode
Executable
Length (Hex)
01
Calibration Mode
Executable
1-byte Type A Data
'Write' Description
(Hex)
Decimal point position (available settings:
0~0x3F)
0x00: No decimal point
GG
0x01: 0.
0x02: 0.0 (one decimal place)
0x04: 0.00 (two decimal places)
0x14: 00.00.00
42
Normal Mode
Not executable
Sub-Command (Hex)
05
Normal Mode
Executable
ZSME000000026