3 . Connect the Device
Run the APP and select [WEIGHT] menu
 Click [Platform no] to enter device list.
 Click the device to be connected.
 Once connected, weight value and device name display on app.
4. Link/Pair LED always on means pairing is successful.
5. Specification Setting
After connecting TM and APP, calibration parameter can be set.
Unit Selection
Max. cap
Division Selection
Decimal Point Selection
Resolution
Max. Show
Cal. Weight
Linearity Segment
Zero Track (D)
Zero Key Range (+/- %)
Initial Zero Range (+/- %)
Calbrating G
Used Area G
6. Linearity Calibration/ Calibration
Run the APP and select [WEIGHT] menu
 After 5. Specification Setting, click [Calibration Start] to start linearity calibration.
TM
Select a unit (mg,g,kg,t,oz,lb)
Entering the maximum capacity
Select division (1,2,5,10)
Select numbers of decimal points
Max. cap/division
No need to enter. Value = Max. cap + 9*division
Entering calibration weight (≤ Max. Capacity)
Select numbers of segments
Entering Zero Tracking range
[ZERO] key range
Entering Initial Zero Range
Entering G value used in verification (default: 9.18914)
Entering local G value (default: 9.18914)
10
Battery: click it to show
ExBLE-P:device name
: click it to disconnect
: click it to show
remaining battery
(Simple weight format)
remaining battery
(EXCELL format)
ZSME000000026