Download Print this page

KYOWA WGA-680A Instruction Manual page 25

Instrumentation amplifier
Hide thumbs Also See for WGA-680A:

Advertisement

Function
Display
updating rate
Display Additional
Function
LEVEL TEST
LEVEL TEST values
signals
LEVEL TEST
ON/OFF
Comparator1 High/Low
Limit assignment
Comparator2 High/Low
Comparator
Limit assignment
functions
Comparator3 High/Low
Limit assignment
Comparator4 High/Low
Limit assignment
9-3. COMMANDS FOR THE CONTROL INPUT SIGNALS
Function
ZERO command ON
Control input
RESET command ON
signals
HOLD command ON
9-4. ERROR CODES
Error message
The measured value is exceeding the
OFL1
input range to the minus side.
The measured value is exceeding the
OFL2
input range to the plus side.
OFL3
The indicated value is below -99999.
OFL4
The indicated value is above +99999.
The sensitivity registering value is
E-00
exceeding the specified range.
E-01
RAM error
E-02
EEPROM error
A read error or checksum error occurs in
E-04
the pre-set data area at startup.
A read error or checksum error is
E-EP
detected in the pre-set data area at startup.
An error occurs while writing the data in
E-E0
the pre-set data area.
An error occurs while writing the data in
E-E1
the calibration data area.
Item
Command
0 9 6 1
0 3 1 1 0: Invalid 1: Valid
0 3 1 2 -99999 to +99999
0 3 1 3 0: LEVEL TEST OFF 1: LEVEL TEST ON
0 5 5 1
0 5 6 1
0 5 7 1
0 5 8 1
Item
Command
0 0 0 0 11
0 0 0 0 13
0 0 0 0 15
Contents
0:15.60 times/sec 1:7.80 times/sec 2:3.90 times/sec
3:1.95 times/sec 4:0.98 times/sec 5:0.49 times/sec
6:0.24 times/sec 7:0.12 times/sec
0: High limit
1: Low limit
0: High limit
1: Low limit
0: High limit
1: Low limit
0: High limit
1: Low limit
Decrease the input signal.
Use the input signal within ±3.2 mV/V.
Decrease the input signal or re-set the sensitivity
registering value.
Use the sensitivity registering value 10000 or less per
0.1 mV/V.
Failure
Contact KYOWA or our representatives for repair.
To clear error messages, press the CANCEL key. When
the error messages are not cleared, the product is
failure.
Contact KYOWA or our representatives for repair.
22
Pre-set value
Pre-set value
Countermeasures