Download Print this page

KYOWA WGA-680A Instruction Manual page 23

Instrumentation amplifier
Hide thumbs Also See for WGA-680A:

Advertisement

8-3-3. Contents of the remote input RX (R → M)
Name
Command processing
response 1
Command processing
response 2
WRITE_READ response
CPU normal operation
Operation pattern 0 to 3
response
Decimal point position 10
3
to 10
[HH]_Comparator3
[HI]_Comparaor1
Working properly_Passed
[LO]_Comparator2
[LL]_Comparator4
Detecting
Holding
OVER error (input)
OVER error (indicated
value)
Calibration error
Memory error
Error status flag
Remote READY
Turn ON when the writing of the pre-set values finished.
Before turning OFF, make sure the [Command processing request 1] is OFF.
Turn ON when the writing of the commands finished.
Before turning OFF, make sure the [Command processing request 2] is OFF.
Become the same status as that of the READ/WRITE signal, at the same timing
when the [Command processing response 2] turns ON.
You are able to check whether the product CPU is working properly by using the
normal operation signal. The CPU normal operation turns ON and OFF every second
when the product is working properly.
Indicate the current operation pattern No.
0
Indicate the decimal point position.
Turn ON when satisfying the setting conditions of the [HH] comparator
(comparator3).
Turn ON when satisfying the setting conditions of the [HI] comparator
(comparator1).
Turn ON when not satisfying the setting conditions of all comparators.
Turn ON when satisfying the setting conditions of the [LO] comparator
(comparator2).
Turn ON when satisfying the setting conditions of the [LL] comparator
(comparator4).
Turn ON during the detecting interval.
Turn ON during the holding interval.
Turn ON when the sensor input is exceeding the input range.
Turn ON when the indicated value is lower than -99999 or is exceeding +99999.
Turn ON when the sensitivity registering value is exceeding the specified range.
Turn ON when the RAM error or EEPROM error was occuuored.
Turn ON when any of the above error was occured.
Turn ON when the initialization is finished and the [Error status flag] was OFF. Turn
OFF when the [Error status flag] turns ON.
Contents
20

Advertisement

loading

This manual is also suitable for:

Wga-680a-04Wga-680a-14