Branson Emerson Ultraseal 20 Operating Manual page 76

Table of Contents

Advertisement

7.2.1.2
Height Calibration
CAUTION
NOTICE
1. Position a 1 mm shim on the tip.
2. From the controller press CALIBRATE. The horn comes down 8 times on the 1 mm shim.
"Calibration Step 1 done" message is displayed.
3. Position a 6 mm shim on the tip.
4. From the controller press CALIBRATE. The horn comes down 8 times on the 6 mm shim.
"Calibration done" message is displayed. If message "Unsuccessful Calibration" is displayed,
repeat steps 1 through 4.
5. Disconnect the RF cable from the actuator.
6. Set the weld mode to Time:
a.
On a VersaGraphiX controller, on the Setup Screen go to Advanced Settings select Time
as the weld mode.
b.
On a Touchscreen controller, go to the Weld Mode screen (MENU>SETTINGS>WELD-
MODE) and select Time as the weld mode.
7. Set the weld time to 0.2 s:
a.
On a VersaGraphiX controller, on the Setup Screen, under Quality Settings press the
button next to the time icon and enter a value of 0.2 s.
b.
On a Touchscreen controller, go to the Weld Settings screen (MENU>SETTINGS) and
press on the Time button and enter a value of 0.2 s.
8. Perform a weld cycle on a 1 mm shim.
9. Adjust height readings to account for tooling variations:
a.
On a VersaGraphiX controller, on the Setup Screen go to Advanced Settings and enter
a value of 1 to the measured height on the right-side column of the Height Off-set.
b.
On a Touchscreen controller, go to the Adjustment screen and enter a value of 1 by
touching the ADJUST button.
10. Connect the RF cable to the actuator.
66
Read all steps completely and exercise caution as tooling moves
during the calibration process.
Readings are consistently better if the calibration is done at 25 PSI.
DCM00051 REV. 10