Figure 4.7 Setting Height Proximity Switch - Branson Emerson Ultraseal 20 Operating Manual

Table of Contents

Advertisement

Figure 4.7
• Turn system on its right side (tooling facing toward you)
• Set/Stack feeler gage to final seal height target as required (recommended starting height target
= 1.0-1.2 times tubing single wall thickness)
• Carefully position feeler gage stack into tooling
• Raise the anvil (step reference- Menu> Maintain> Anvil) to contact feeler gage stack
• Loosen lock set screw on side of bottom plate
• Carefully adjust prox switch screw on bottom plate until height switch "Red Light" illuminates
• Tighten prox switch lock set screw on side of bottom plate
• Lower the anvil (step reference- Anvil)
• Verify that setting is correct via a GO/NO GO check
• "GO" = Feeler gage height (1.0-1.2 times tubing single wall thickness) – (.05mm)
• "NO GO" = Feeler gage height (1.0-1.2 times tubing single wall thickness) + (.05mm)
• Carefully position "GO" gage in tooling
• Raise the anvil (step reference- Anvil) to contact feeler gage stack
• Height switch (Red light) should illuminate. If not, lower the anvil (step reference- Anvil), remove
"GO" gage and repeat setting procedure
• Lower the anvil (step reference- Anvil)
• Carefully position "NO GO" gage in tooling
• Raise the anvil (step reference- Anvil) to contact feeler gage stack
• Height switch (Red light) should not illuminate. If illuminated, lower the anvil (step reference-
Anvil), remove "NO GO" gage and repeat setting procedure
• Lower the anvil (step reference- Anvil)
• Store all gages
• Turn system upright
• Return to run mode (step reference- Exit> Exit)
Produces samples, evaluate sealed tube final height and test as required. Repeat this
procedure as necessary.