Pfeiffer Vacuum CCR 351 Operating Instructions Manual page 18

Ceramic capacitance gauge
Hide thumbs Also See for CCR 351:
Table of Contents

Advertisement

Evacuate the gauge to a pressure according to the
following table:
If the final pressure is too high for a zero adjustment (>25%
of F.S.), zero cannot be reached.
Briefly press the button with a pin (max. ø1.1 mm), or ...
... apply supply voltage to Remote Zero at pin 7 (Pulse
→  7).
Zero adjustment runs automatically (duration ≈ 8 s).
18
F.S.
<5×10
1000 Torr
100 Torr
<5×10
<5×10
10 Torr
max. ø1.1 mm
After the zero adjustment, the gauge automatically
returns to the measurement mode.
Recommended final pressure
for zero adjustment
-2
Torr
<7×10
-3
Torr
<7×10
-4
Torr
<7×10
Taster kurz drücken
Press the button briefly
-2
-2
hPa
<7×10
-3
-3
hPa
<7×10
-4
-4
hPa
<7×10
BG 5140 BEN
(2017-09)
mbar
mbar
mbar