Datascope Accutorr Plus Service Manual page 36

Hide thumbs Also See for Accutorr Plus:
Table of Contents

Advertisement

Alarm Limit Table
PARAMETER
RANGE
UNITS
FACTORY
UNITS OF
DEFAULT
INCREMENT
Systolic High
Adult
Off, 60-260
Pediatric
Off, 60-160
mmHg
Off
5
Neonate
Off, 50-125
Systolic Low
Aduit
Off, 55-150
Pediatric
Off, 55-130
mmHg
Off
5
Neonate
Off, 45-115
Diastolic High
Adult
Off, 40-200
Pediatric
Off, 40-150
mmHg
Off
5
Neonate
Off, 35-100
Diastolic Low
Adult
Off, 30-120
Pediatric
Off, 30-50
mmHg
Off
5
Neonate
Off, 25-50
MAP High
Adult
Off, 90-200
Pediatric
Off, 90-150
mmHg
Of
5
Neonate
Off, 60-110
MAP
Low
Adult
Off, 40-100
Pediatric
Off, 40-70
mmHg
Of
5
Neonare
Off, 30-70
Pulse Rate High
Adult
Off, 100-245
Pediatric
Off, 100-245
bpm
Off
5
Neonate
Off, 100-245
Pulse Rate Low
Adult
Off, 35-120
Pediatric
Off, 35-150
bpm
Of
5
Neonate
Off, 75-200
SpO2 High
Adult
Off, 61-99
Pediatric
Off, 61-99
%SpO2
Of
1
Neonate
OFf,61-99
SpO2 Low
Adult
60-95
Pediatric
60-95
%SpO2
86
1
Neonate
60-95
1.3.5.2
Alarm Violations
An alarm condition exists if the parameter is equal to or is outside the high/low limit
range that has been set. When an alarm limit is violated, the following actions occur:
e The LEDs for the parameter in an alarm condition flashes.
e The parameter in an alarm condition is in reverse video on the LCD (Accutorr Plus models
with Trend Screen).
« The alarm tone is sounded (unless muted with the MUTE key (30)).
e The parameter(s) chat was in an alarm condition will be in brackets [ ] when printed on
the recorder.
1-26
Accutorr Plus Service Manual
Chapter 1, Operation