Nibp Pressure Limit Fail Safe; Cuff Inflation Time; Automatic Adjustment Of Cuff Inflation Pressure - Datascope Accutorr Plus Service Manual

Hide thumbs Also See for Accutorr Plus:
Table of Contents

Advertisement

1.3.3.1
NIBP Pressure Limit Fail Safe
If the cuff is over-pressurized, it will automatically deflate and the status code 8812
(STOP - CUFF OVERPRESSURE) ot error code 987 (STOP - HARDWARE
OVERPRESSURE) will be displayed in the Room/Bed or display.
The unit must be turned off and back on again to reset the hardware
overpressure switch (error code 987) before any new measurements can be taken.
1.3.3.2 Cuff Inflation Time
If the cuff pressure does not attain 20 mmHg within 40 seconds of the start of inflation
or if the target pressure is not reached within another 60 seconds, then the cuff is deflated
and status codes will be displayed in the Room/Bed display. See section 1.3.16 for a list of
error and status codes.
-
1.3.3.3 Automatic Adjustment of Cuff Inflation Pressure (Adaptive
Inflation)
The unit adjusts the inflation pressure according to the previous reading of the systolic
pressure. After the first successful measurement, the inflation pressure is the previous
systolic +50 mmHg in the adult mode and +50 mmHg in the pediatric mode and +40
mmHg in the neonate mode.
To view the current inflation pressure, press and hold (2 beep tones, approximately 3
seconds) the Patient Setup Key (37). The current inflation pressure is shown in the MAP
display. If required, use the Patient Info. Up & Down Arrow keys (27 & 28) to change
the inflation pressure.
It is also possible to permanently override this adjustment in the User Configuration. See
section 1.3.15 for details.
1-22
Accutorr Plus Service Manual
Chapter 1, Operation