6 Operation
6.3 Controlled shutdown and removal of flow kit
Alternative 2 - Drain and remove
the flow kit by section
140
TeDIS Valid on Date Printed
Step
Action
3
Disconnect the flow kit sensors from the system.
4
Make sure to fit caps to all sensor connections on the system with the caps
provided on the system.
5
Disconnect and remove customer tubing.
6
Set the switch to Flowkit install to open the pinch valves, and press Re-
set.
7
Remove the flow kit sections starting from the outlet/column side by
opening the pinch valves locking pins one by one as the tubing is removed
from the valve. Close each valve after the tubing is removed.
8
Dispose of the filled flow kit and column in accordance with local regula-
tions.
To drain the flow kit, and then remove each section of the flow kit follow the instruc-
tions below.
NOTICE
The inlets and pump sections are not entirely drained using this
procedure, therefore splashing is to be expected.
Step
Action
1
Connect a drain tube to the Column bottom outlet tubing and lead it to a
suitable waste container.
2
Close any valves to tanks leading to the system inlets.
3
Set the switch to Drain, and press Reset.
4
When no more waste is noticed, disconnect and remove customer tubing
from the Column bottom outlet.