Equipment Heat Output (Dissipation) - GE Innova 2000 Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Medical Systems
REV 5
1-2

Equipment Heat Output (Dissipation)

Refer to Table 15. To obtain heat output information for components not specified inTable 15, refer to
the appropriate component Pre–Installation Manual listed in Chapter 2.
Table 15 – INNOVA 2000 SYSTEM HEAT OUTPUTS (BY COMPONENT)
PRODUCT OR COM-
PRODUCT OR COM
PONENT
Innova LC Positioner
POSTIT
Cabinet
Omega
Patient table
Maxiray 150 V5
X–Ray tube assy
Innova Digital Detector
ACAB
Cabinet
VMP
Cabinets
Tube Chiller
Digital Detector Chiller
Innova 2000
Operator console
DL user area
Note 1: Air intake requirements 1360 cubic meter per hour
2: For more details, consult appropriate pre–installation manual
Innova 2000 Cardiovascular Imaging System
HEAT OUTPUT (BTU/HR)
STANDBY
IN–USE
negligible
5600
negligible
negligible
3400
5120
20470(1/2)
181
33
HEAT OUTPUT (WATTS)
STANDBY
1166
negligible
11123
1640
negligible
1,000
negligible
5100
1000
1500
194
1000
Pre–Installation Manual
pim 2337741–100
IN–USE
342
3260
300
2500/2800W
1500
6000W
250W
53
57
300