3.2.1 Atmospheric pressure

Ensure that the in-use and storage locations meet the atmospheric pressure minimums and
maximums listed in
Table 3-4 Atmospheric pressure requirements
70.1 kPa
Ensure that the in-use and storage location conditions do not fluctuate more than the maximum
rates of change given in
Table 3-5 Atmospheric pressure change rate limits

Section 3.3 Heat output

Below are the maximum power and thermal values in summary form:
Table 3-6 Optima XR200amx/XR220amx heat output
Charge + Expose
For reference, a tightly closed room of 3 m x 3 m (10 ft x 10 ft) with a 2.4 m (8 ft) ceiling would require
2100 BTU/hour to raise the temperature by 12.2°C (10°F).

Section 3.4 Audible noise

Noise limits assume that all system covers are installed, and assume background noise of 40 dBA.
Measurements are taken 1 m (3.3 ft) from the equipment. The in-use measurement is taken when
the unit is being driven. Tube rotor and exposure tone noise are not included in these
measurements.
Maximum operating noise generated by the equipment is listed in
Table 3-7 Audible noise produced by the system
Upper limit: 55 dBA
Page 28
3
EVISION
Table
3-4.
In-use
Min.
Max.
106 kPa
Table
3-5.
In-use
< 1.8 kPa/hour
State
Watts
Idle
143
Drive
275
Charge
261
Expose
474
611
In-use
Section 3.3 Heat output
Storage
Min.
Max.
70.1 kPa
106 kPa
Storage
< 76 kPa/hour
BTU/hour
488
938
890
1616
2084
Idle
Upper limit: 50 dBA
O
XR200
/XR220
PTIMA
AMX
P
-I
M
RE
NSTALLATION
Table
3-7.