Joerns® Camtec Series Bed Frames
RC 750 Bed

Assembly

Disassembly of the split frame for mobility
Warning: Lock casters when cleaning, inspecting or performing any maintenance on the bed. Lower all mattress
support platform sections unless necessary for service access.
1. When transferring the patient into or out of the bed all four casters should be in the locked position.
2. Disconnect the power cord from the wall outlet. Lock all casters.
3. Remove the footboard from the outward most sockets. If this is a scale model, disconnect the two leads from the
bed to the readout.
4. Remove the headboard from the inward most sockets. The power cord may be removed from the power cord
hanger as needed.
5. The pendant should remain with the head end of bed. If it becomes necessary to unplug the pendant, squeeze
the release levers on either side of the plug, and then gently remove the plug.
6. Disconnect the electrical cables between the two halves at the center point of the frame. A set of red-banded
cables and the black cables will be disconnected.
7. Loosen and remove the ⅜" and ½" bolts in both sides of the bed frame (Figure 1). Raise the coupling end of the
head end and foot end. Separate the two halves as indicated.
8. Transfer the head and foot end to their new location. Keep the head and foot section matched together per serial
number ID labels. Transfer the head board and foot board to their new location.
Assembly of the split frame once in position
9. Place the head and foot end in their approximate position. Note the head and foot sections are to be used with
matching serial number ID labels.
10. Raise the coupling end of the head end and foot end (Figure 1). Couple the two halves together as indicated.
Install the ⅜" and ½" bolts in both sides of the bed frame. Tighten nuts and bolts with wrenches.
11. Connect the electrical cables between the two halves at the center point of the frame. Plug the red-banded
cables into each other and the black cables into each other. Make sure cable plugs are properly seated and
locked together.
12. If the pendant is not already
attached to the bed, plug the hand
pendant into the control module.
Make sure the pendant plug is
securely fastened in the connector,
it should snap in place. Fasten
the pendant cord to the frame
using the cable clamp provided.
If it becomes necessary to unplug
the pendant, squeeze the release
levers on either side of the plug,
then gently remove the plug.
13. Install the headboard posts in the
inward most sockets provided
at the head end with the board
closest to the mattress. The power
cord hanger should be on the side
away from the mattress.