  • Page 1 User-Service Manual Joerns Bed Frames ® WeCare™ Bed To avoid injury, read user’s manual before using.
  • Page 2: Important Precautions

    Prior to use, ensure the wall outlet. Injury may result from improper routing of the therapeutic benefits outweigh the risk of entrapment. © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 3 (for which CISPR 11 class B is normally required) this ® specifically for the WeCare bed and it is locked in place. equipment might not offer adequate protection to radio- Failure to lock assist devices may result in injury or death.
  • Page 4: Table Of Contents

    That is why Joerns offers a wide array of compliance Advanced ergonomic design will provide years of options. We assist customers in selecting compliant reliable service within your care environment.
  • Page 5: Features And Specifications

    The identification label is Class II Equipment located on bed frame under the mattress support platform foot section. Protective Earth (ground) Pinch Point © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 6: Assembly

    5. Insert safety pin on left and right sides, 2 pins total (Figure 5). Be sure to route the actuator cables through the nearest safety pin bail (Figure 6). Figure 5 Figure 1 Figure 6 © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 7 10. Route power cord in grey clips to head end. Plug-in power cord to power source and raise the bed frame. Figure 10 Note: Watch for obstructions. Figure 11 Figure 7 Figure 12 Figure 8 Figure 9 Figure 13 © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 8 WeCare™ Bed Platform has four IV sockets ® as standard equipment. There are two sockets at head end of bed and two at foot end (Figure 18). The IV sockets are used with Joerns IV poles and most ½” IV Socket IV poles. Figure 18...
  • Page 9 (Figure 21). 2. Tighten all screws with panels in place. 3. Panels should slip in and out of their sockets easily and securely. Figure 21 © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 10 (Figure 24). 6. Zip tie as shown by arrows (Figure 25). ARP NOTE: Need high res image from Gary’s PPT. Figure 22 Figure 25 Figure 23 © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 11: Operation

    Mobile Locked Position Swivel Casters Position Any Height (casters Head & Foot (in Low only must be locked in Low) Low position only) Figure 28 © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 12: Servicing

    Overload protection is achieved via current sensing technology, which will automatically reset. If any bed function fails to operate, check to see if bed is overloaded or constrained from movement. Figure 31 © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 13: Troubleshooting

    AC inlet receptacle on the control box. Simultaneously, pull gently outward on AC power cord where it connects to control box until plug comes out. © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 14: Cleaning

    Clean all electrical components (actuators, control box, pendant and cables) on your Joerns bed with a damp, ® well-wrung cloth. Excess water in or on electrical parts can damage these components. © 2022 Joerns Healthcare • 6110435 RevC • 22-2083...
  • Page 15: Accessories

    WeCare™ Bed, particularly when it is lowered 1. We recommend using the Joerns trapeze with the ® onto its casters. Note: The WeCare Bed is not Joerns trapeze adapter. designed for use as resident/patient transport 2. Install the trapeze assembly according to devices.
  • Page 16 Some trapeze systems may not be holder at that height are especially at risk. Failure to do suitable for use with the WeCare Bed, due to its so could result in injury or death. low operating height.
  • Page 17 Manufacturer’s Declaration –Electromagnetic Emissions The Joerns WECC bed is intended for use in the electromagnetic environment (for professional healthcare) specified below. The customer or the user of the Joerns WECC bed should assure that it is used in such an environment.
  • Page 18 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity: The Joerns WECC bed is intended for use in the electromagnetic environment (for professional healthcare) specified below. The customer or the user of the Joerns WECC bed should assure that it is used in such an environment. Immunity Test...
  • Page 19 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The Joerns WECC bed is intended for use in the electromagnetic environment (for professional healthcare) specified below. The customer or the user of the Joerns WECC bed should assure that it is used in such and environment.
  • Page 20: Warranty

    +44 (0)844 811 1156 +31 (0)30 6363799 Charlotte, NC 28273 (P) 800.966.6662 (P) 866.546.1151 +44 (0)844 811 1157 (P) 800.826.0270 (F) 800.232.9796 (F) 519.451.8662 (F) 800.457.8827 www.joerns.com • email: info@joerns.com © 2022 Joerns Healthcare • 6110435 RevC • 22-2083 • MAN1...