Error Code Table - Eizo Rad i Force MX215 Instructions For Use Manual

Color lcd monitor
Table of Contents

Advertisement

Error code table

Error Code
0***
1***
*1**
*2**
*3**
**10
**11
**34
**61
**95
• Errors that occurred during SelfCalibration
• Errors that occurred during Grayscale Check
• Errors that occurred during DICOM
• Errors that occurred during Custom
• Errors that occurred during CAL
• The product maximum brightness may be lower than the target brightness.
• Lower the target brightness.
• The product minimum brightness may be higher than the target brightness.
• Raise the target brightness.
• The sensor may not have come out during calibration, or light may have entered the
sensor.
• Turn off the main power, wait for a few minutes before turning the power back on, and
then execute SelfCalibration/Grayscale Check again.
• The sensor may not have come out.
• Check whether there is any foreign object near the sensor.
• Execute SelfCalibration/Grayscale Check again.
• The execution conditions for SelfCalibration/Grayscale Check may not be appropriate.
• Check the execution conditions and reset them if necessary.
Description
Chapter 6 Troubleshooting
33