Technical Descriptions
Electromagnetic emissions
The RadiForce series is intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Emission test
Compliance
RF emissions
Group 1
CISPR11 / EN55011
RF emissions
Class B
CISPR11 / EN55011
Harmonic emissions
Class D
IEC / EN61000-3-2
Voltage fluctuations /
Complies
flicker emissions
IEC / EN61000-3-3
Electromagnetic immunity
The RadiForce series has been tested at the following compliance levels according to the testing requirements for
professional healthcare facility environments defined in IEC / EN60601-1-2.
Customers and users of the RadiForce series must ensure that the RadiForce series is used in the following
environments:
Immunity test
healthcare facility
Electrostatic
±8 kV contact discharge
discharge (ESD)
±15 kV air discharge
IEC / EN61000-4-2
Electrical fast
±2 kV power lines
transients / bursts
±1 kV input / output lines
IEC / EN61000-4-4
Surges
±1 kV line to line
IEC / EN61000-4-5
±2 kV line to ground
Voltage dips, short
0 % U
interruptions and
0.5 cycles and 1 cycle
voltage variations on
70 % U
power supply input
25 cycles
lines
0 % U
IEC / EN61000-4-11
5 sec
Power frequency
30 A/m
magnetic fields
(50 / 60 Hz)
IEC / EN61000-4-8
48
Appendix
The RadiForce series uses RF energy only for its internal function.
Therefore, its RF emission are very low and are not likely to cause any interference in
nearby electronic equipment.
The RadiForce series is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Test level for
Compliance level
professional
environments
±8 kV contact discharge
±15 kV air discharge
±2 kV power lines
±1 kV input / output lines
±1 kV line to line
±2 kV line to ground
(100 % dip in U
)
0 % U
T
T
0.5 cycles and 1 cycle
(30 % dip in U
)
70 % U
T
T
25 cycles
(100 % dip in U
)
0 % U
T
T
5 sec
30 A/m
Electromagnetic environment - Guidance
Electromagnetic environment -
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
(100 % dip in U
)
Mains power quality should be that of a typical
T
T
commercial or hospital environment. If the user
(30 % dip in U
)
of the RadiForce series requires continued
T
T
operation during power mains interruptions, it
(100 % dip in U
)
is recommended that the RadiForce series be
T
T
powered from an uninterruptible power supply
or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment. The
product should be kept at least 15 cm away
from the source of power frequency magnetic
fields during use.