Page 1
Instructions for Use ■User’s Manual Important Please read PRECAUTIONS, this User’s Manual, and the Setup Manual (separate volume) carefully to familiarize yourself with safe and effective usage. • Please refer to the Setup Manual for basic information ranging from connection of the monitor to a PC to using the monitor. •...
Page 2
EIZO Corporation. EIZO Corporation is under no obligation to hold any submitted material or information confidential unless prior arrangements are made pursuant to EIZO Corporation’s receipt of said information. Although every effort has been made to ensure that this manual provides up-to-date information, please note that EIZO monitor specifications are subject to change without notice.
The backlight of the LCD panel has a fixed lifetime. When the screen becomes dark or begins to flicker, please contact your local EIZO representative. Do not press on the panel or edge of the frame strongly, as this may result in display malfunctions, such as interference patterns, etc.
Cleaning Attention • Chemicals such as alcohol and antiseptic solution may cause gloss variation, tarnishing, and fading of the cabinet or panel, and also quality deterioration of the image. • Never use any thinner, benzene, wax, and abrasive cleaner, which may damage the cabinet or panel. Note •...
Chapter 1 Introduction Thank you very much for choosing an EIZO color LCD monitor. 1-1. Features • 21.3 inches • Supports a resolution of 2M pixels (Portrait: 1200 × 1600 dots (H × V)) • Uses high-contrast panel (1500:1). Enables the display of sharp images.
1-2. EIZO LCD Utility Disk An “EIZO LCD Utility Disk” (CD-ROM) is supplied with this product. The following table shows the disk contents and the overview of the software programs. ● Disk contents and software overview The disk includes application software programs for adjustment, and User’s Manual. Refer to Readme.txt file on the disk for software startup procedures or file access procedures.
1-3. Basic Operations and Functions ● Basic Adjustment menu operations Displaying Adjustment menu 1. Press . The Adjustment menu appears. Menu title Current mode Item Setting Menu Adjusting/Setting 1. Select a menu to adjust/set with or , and then press 2.
● Functions The following table shows the Adjustment menu and setting items of each menu. Main Menu Item Reference Color “2-3. Adjusting Color” (page 13) Brightness Temperature Gamma Advanced Settings Saturation Gain “3-7. Resetting to Default” (page 23) Color Reset RadiCS SelfQC “4-1.
Page 11
Main Menu Item Reference Optional Settings Logo “3-5. Showing and hiding the EIZO logo” (page 22) Key Lock “3-6. Locking Operation Buttons” (page 23) Forced Analog “4. No picture when using analog signal input” (page 31) DP Power Save “5-3. Setting DisplayPort Power Save” (page 28) Orientation “3-1.
Chapter 2 Adjusting Screens 2-1. Compatible Resolutions For details on compatible resolutions, refer to “Compatible Resolutions” in the Setup Manual. 2-2. Setting the Resolution ● Windows 8 / Windows 7 1. For Windows 8, click the “Desktop” tile on the Start Screen to display the desktop. 2.
2-3. Adjusting Color ● Selecting the display mode (CAL Switch mode) A desired mode can be selected easily according to the monitor application. Display Modes √: Calibration is available. Mode Purpose 1-DICOM Available to display in DICOM mode. √ 2-Custom Available for the setting according to your preference.
● Adjustable settings in each mode The adjustable settings depend on the mode. (Non-adjustable or non-settable function cannot be selected.) √: Adjustable -: Not adjustable CAL Switch mode Icon Function 1-DICOM 2-Custom 3-CAL 4-Text Brightness √ √ √ √ Temperature √...
● Adjusting Temperature The color temperature can be adjusted. The color temperature is normally used to express the hue of “white” and/or “black” by a numerical value. The value is expressed in degrees “K” (Kelvin). The screen becomes reddish at a low color temperature, and bluish at a high color temperature, like the flame temperature.
● Adjusting Saturation The color saturation can be adjusted. Setting Range -100 to 100 Procedure 1. Select “Color” from the Adjustment menu, and press 2. Select “Advanced Settings” from “Color”, and press 3. Select “Saturation”, and press 4. Use for adjustment. 5.
2-4. Adjusting Screens ● Digital Signal Input When digital signals are input, images are displayed correctly based on the preset data of the monitor, but if you perform more advanced adjustment, see “2-3. Adjusting Color” (page 13) and subsequent pages. ●...
Page 18
Perform the auto adjustment again with the analog screen adjustment pattern displayed. ● To adjust flickering, screen position, and screen size automatically Procedure 1. Display Pattern 1 in full screen on the monitor using the “Screen adjustment pattern files”. 2. Choose “Screen” from the Adjustment menu, and press 3.
Page 19
● To remove flickering or blurring Procedure 1. Choose “Analog Adjustment” from “Screen”, and press 2. Choose “Phase” from “Analog Adjustment”, and press The “Phase” menu appears. 3. Adjust the phase with to minimize flickering or blurring. 4. Press to exit. When vertical bars appear on the screen after adjustment, go back to “To eliminate vertical bars”. (Clock →...
Adjusting color gradation. ● To adjust color gradation automatically Every color gradation (0 to 255) can be displayed by adjusting the signal output level. Procedure 1. Display Pattern 2 in full screen on the monitor using the “Screen adjustment pattern files”. 2.
Chapter 3 Setting Monitor 3-1. Setting Orientation This function changes the display orientation of the image based on the operating environment. Attention • Refer to the User’s Manual of the graphics board. • Setting is available even when the PC is not started up. Procedure 1.
3. Select the position with or . 4. Press to exit. 3-5. Showing and hiding the EIZO logo You can choose to show or hide the display of the EIZO logo. Procedure 1. Press to turn off the monitor. 2. Hold down while pressing for at least two seconds to turn on the monitor.
3-6. Locking Operation Buttons The settings can be locked to prevent them being changed. Procedure 1. Press to turn off the monitor. 2. Hold down while pressing for at least two seconds to turn on the monitor. The “Optional Settings” menu appears. 3.
Chapter 4 Controlling Monitor Quality This product has the built-in Integrated Front Sensor, which allows the user to perform periodic calibration and Grayscale Check for the monitor independently. Attention • When a panel protector is attached, the Integrated Front Sensor cannot be used. Do not execute “SelfCalibration” or “Grayscale Check”...
● Viewing SelfCalibration result Attention • The result can be viewed when “QC History” is “On” (see “4-4. Setting On/Off for Warning and QC History” (page 26)). Procedure 1. Select “RadiCS SelfQC” from the Adjustment menu, and press 2. Select “SelfCalibration” from “RadiCS SelfQC”, and press 3.
4-3. Acquiring Life Prediction Data The data necessary to predict the monitor life can be acquired. Attention • When a panel protector is attached, be sure to set “Meas. Frequency” to “Off”. Note Upon reaching the timing, the sensor will activate and start measurement. •...
Chapter 5 Power Saving Functions 5-1. Setting Power Save The monitor can be set to the power saving mode according to the PC status. When the monitor has shifted to the power saving mode, the images on the screen are not displayed. Attention •...
5-3. Setting DisplayPort Power Save The power setting of the DisplayPort of the monitor can be switched in the power saving mode. Attention • When “On” is set, the display position on the window may not be positioned correctly when the monitor power is turned on or off.
5-4. Setting the Power Saving Mode when a Person Leaves the Monitor The sensor on the front of the monitor detects the movement of the operator when Presence Sensor is set to “On”. When a person moves away from the monitor, it shifts to the power saving mode automatically and does not display the images on the screen to reduce the power consumption.
Chapter 6 Troubleshooting If a problem still remains after applying the suggested remedies, contact your local EIZO representative. 6-1. No Picture Problem Possible cause and remedy 1. No picture • Check whether the power cord is connected properly. • Power indicator does not light.
• Use “Brightness” in the Adjustment menu to adjust it. (The LCD monitor backlight has a limited life span. When the screen becomes dark or begins to flicker, contact your local EIZO representative.) 2. Characters are blurred. • Check whether the PC is configured to meet the resolution and vertical scan frequency requirements of the monitor (see “Compatible Resolutions”...
• Change the USB port to another one. If the PC or peripheral devices works correctly by changing the USB port, contact your local EIZO representative. (Refer to the manual of the PC for details.) • Reboot the PC.
● Error code table Error Code Description 0*** • Errors that occurred during SelfCalibration 1*** • Errors that occurred during Grayscale Check *1** • Errors that occurred during DICOM *2** • Errors that occurred during Custom *3** • Errors that occurred during CAL **10 •...
Chapter 7 Reference 7-1. Attaching Arm An optional arm (or an optional stand) can be attached by removing the stand section. Please refer to our web site for the corresponding optional arm (or optional stand). http://www.eizoglobal.com Attention • When attaching an arm or stand, follow the instructions of their User’s Manual. •...
7-2. Connecting Multiple PCs The product has multiple connections to PCs and allows you to switch to one of the connections for display. Connection examples DisplayPort DVI-I connector connector Digital Digital (DVI) (DisplayPort) Signal cable Signal cable DVI-D - DVI-D DisplayPort - DisplayPort (supplied) (supplied)
● Setting input signal switching procedure Setting Function Auto The monitor recognizes the connector through which PC signals are input. When a PC is turned off or enters the power saving mode, the monitor automatically displays another signal. Manual The monitor displays the signal from the selected connector despite the signal input. Select an active input signal with Procedure 1.
7-3. Making Use of USB (Universal Serial Bus) This monitor has a USB hub. It works as a USB hub when connected to a USB-compatible PC, allowing the connection of peripheral USB devices. ● Required system environment • A PC equipped with a USB port •...
7-4. Showing Monitor Information ● Displaying signal information The information about the current input signals can be displayed. Procedure 1. Select “Tools” from the Adjustment menu, and press 2. Select “Signal Info” from “Tools”, and press The “Signal Info” appears. (Example) ●...
7-5. Specifications LCD panel Size 21.3 inch (540 mm) Type Color TFT, LED backlight Surface treatment Anti-glare Surface hardness Viewing angles Horizontal 178˚, vertical 178˚ (CR≥10) Dot pitch 0.270 mm Response time Black-white-black: Approx. 20 ms Horizontal scan Analog 26 - 100 kHz frequency Digital 31 - 100 kHz...
Page 41
209.5 245.5 HIGHEST POSITION Accessories Calibration Kit EIZO “RadiCS UX1” Ver. 4.1.4 or later EIZO “RadiCS Version Up Kit” Ver. 4.1.4 or later Network QC Management Software EIZO “RadiNET Pro” Ver. 4.1.4 or later Cleaning Kit EIZO “ScreenCleaner” Panel protector FP-2101 *1 When a panel protector is attached, the Integrated Front Sensor cannot be used.
7-6. Preset Timing The following table shows factory preset analog signal timing. Attention • Display position may be deviated depending on the PC connected, which may require screen adjustment using Setting menu. • If a signal other than those listed in the table is input, adjust the screen using the Setting menu. However, screen display may still be incorrect even after the adjustment.
Chapter 8 Glossary Clock The analog signal input monitor needs to reproduce a clock of the same frequency as the dot clock of the graphics system in use, when the analog signal is converted to a digital signal for image display. This is called clock adjustment.
Page 44
HDCP (High-bandwidth Digital Content Protection) Digital signal coding system developed to copy-protect the digital contents, such as video, music, etc. This helps to transmit the digital contents safely by coding the digital contents sent via the DVI or HDMI connector on the output side and decoding them on the input side. Any digital contents cannot be reproduced if both of the equipments on the output and input sides are not applicable to HDCP system.
EIZO, the EIZO Logo, ColorEdge, DuraVision, FlexScan, FORIS, RadiCS, RadiForce, RadiNET, Raptor, and ScreenManager are registered trademarks of EIZO Corporation in Japan and other countries. ColorNavigator, EcoView NET, EIZO EasyPIX, EIZO ScreenSlicer, i • Sound, Screen Administrator, and UniColor Pro are trademarks of EIZO Corporation.
Medical Standards • It shall be assured that the final system is in compliance to IEC60601-1-1 requirement. • Power supplied equipment can emit electromagnetic waves, that could influence, limit or result in malfunction of the monitor. Install the equipment in a controlled environment, where such effects are avoided.
Anyone who connects additional equipment to the signal input part or signal output parts, con- figuring a medical system, is responsible that the system complies with the requirements of IEC/ EN60601-1-2. Be sure to use the cables attached to the product, or cables specified by EIZO. Use of cables other than those specified or provided by EIZO of this equipment could result in in- creased electromagnetic emissions or decreased electromagnetic immunity of this equipment and improper operation. Cables EIZO Designated Max. Shielding Ferrite Core Cables Cable Length Signal cable PP300 / PP200...
Page 48
Technical Descriptions Electromagnetic emissions The RadiForce series is intended for use in the electromagnetic environment specified below. The customer or the user of the RadiForce series should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment - Guidance RF emissions Group 1 The RadiForce series uses RF energy only for its internal function.
Page 49
Electromagnetic immunity The RadiForce series has been tested at the following compliance levels according to the testing requirements for professional healthcare facility environments defined in IEC / EN60601-1-2. Customers and users of the RadiForce series must ensure that the RadiForce series is used in the following environments: Immunity test Test level for...
Page 50
Recommended separation distances between portable or mobile RF communication equipment and the RadiForce Series The RadiForce series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the RadiForce series can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the RadiForce series.
Page 51
The RadiForce series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. For other portable and mobile RF communication equipments (transmitters), minimum distance between portable and mobile RF communications equipment (transmitters) and the RadiForce series as recommended below, according to the maximum output power of the communications equipment.
Changes or modifi cations not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Note Use the attached specifi ed cable below or EIZO signal cable with this monitor so as to keep interference within the limits of a Class B digital device. - AC Cord...
Page 53
Declaration of compliance with India RoHS We, EIZO Corporation, hereby declare and guarantee that this product has been designed and manufactured in compliance with the E-Waste management rule 2016 which prohibit the inclusion of the following substances except for the exemptions listed in schedule II.
Page 54
Purchaser or any third parties other than as provided under this Warranty. EIZO and Distributors will cease to hold or store any parts of the Product upon expiration of seven (7) years after the production of such parts is discontinued. In repairing the monitor, EIZO and Distributors will use renewal parts which comply with our QC standards.
Page 55
Produkts keinerlei Haftung oder Verpfl ichtung dem Erstkäufer oder Dritten gegenüber. EIZO und die Vertriebe verpfl ichten sich, Ersatzteile für das Produkt über einen Zeitraum von sieben (7) Jahren nach Einstellung der Produktion der Ersatzteile zu lagern bzw. anzubieten. EIZO und seine Vertriebspartner verpfl ichten sich, bei einer etwaigen Reparatur des Monitors ausschließlich Produkte gemäß...
Page 56
EIZO et ses Distributeurs cesseront de tenir ou conserver en stock toute pièce de ce Produit après l’expiration de sstandards zu la période de sept (7) ans suivant l’arrêt de la production de telles pièces. Pour réparer le moniteur, EIZO et ses distributeurs utiliseront des pièces de rechange conformes à nos normes de contrôle qualité.
Page 57
(en lo sucesivo el “Comprador original”) que compró el producto especifi cado en este documento (en riferimento da lo sucesivo el “Producto”) a EIZO o a sus Distribuidores, que EIZO y sus Distribuidores, a su propio criterio, repararán originale”) ch o sustituirán el Producto de forma gratuita si el Comprador original detecta dentro del periodo de la Garantía (indicado...
Page 58
GARANZIA LIMITATA EIZO Corporation (a cui si farà riferimento da qui in poi con “EIZO”) ed i Distributori autorizzati da EIZO (a cui si farà riferimento da qui in poi con “Distributori”) garantiscono, secondo i termini di questa garanzia limitata (a cui si farà...
Page 59
Garantin är endast giltig i de länder där det finns Distributörer. Garantin begränsar inte några av den Ursprunglige köparens lagstadgade rättigheter. Oavsett andra villkor i denna garanti ska inte EIZO Distributörer under några villkor ha något ansvar i något av de fall som beskrivs nedan: (a) När någon bristfällighet hos Produkten kan härledas till att ha uppstått vid frakt, modifiering, ändring, felaktigt...
Page 61
рвоначальному покупателю или по отношению к любым третьим сторонам, кроме обязательств, оговоренных в этой Гара нтии. EIZO и Дистрибуторы прекратят хранение или складирование любых деталей Продукта по истечении семи (7) лет после п рекращения производства таких деталей. В случае ремонта монитора EIZO и Дистрибьюторы будут использовать запчас...
Page 62
Cuando desee deshacerse de este producto, póngase en contacto con un distribuidor o una fi lial de su país. Encontrará las direcciones de contacto en el sitio web de EIZO que se indica a continuación. http://www.eizoglobal.com Informazioni sul riciclaggio Per lo smaltimento e il riciclaggio del presente prodotto, attenersi alle normative vigenti nel proprio paese per ridurre l’impatto ambientale.
Page 63
Ten produkt po zużyciu powinien być zbierany i przetwarzany zgodnie z krajowymi przepisami dotyczącymi ochrony środowiska. Wyrzucając ten produkt, należy skontaktować się z lokalnym dystrybutorem lub partnerem. Adresy kontaktowe można znaleźć we wskazanej poniżej witrynie internetowej fi rmy EIZO. http://www.eizoglobal.com...
Page 64
Při likvidaci produktu musí být produkt vyzvednut a recyklován podle zákonů příslušné země, aby nedocházelo k zatěžování životního prostředí. Zbavujete-li se produktu, kontaktujte distributora nebo pobočku ve své zemi. Kontaktní adresy jsou uvedeny na následující webové stránce společnosti EIZO. http://www.eizoglobal.com Ringlussevõtu alane teave...
Page 65
вашата страна за да се намали замърсяването на околната среда. Когато искате да се освободите от този продукт, моля свържете се с търговския му представител или със съответните органи отговарящи за това във вашата страна. Данните за връзка с нас са описани на следния Интернет сайт на EIZO: http://www.eizoglobal.com Informaţie referitoare la reciclare Acest produs, când debarasat, trebuie colectat şi reciclat conform legislaţiei ţării rspective ca să...
Need help?
Do you have a question about the Rad i Force MX215 and is the answer not in the manual?
Questions and answers