SelfCalibration / Grayscale Check fails.
• Refer to

Error Code Table

• If an error code that does not appear in the error code table is displayed, contact your
dealer or local EIZO representative.
Attention
• Do not touch the Integrated Front Sensor.
The SelfCalibration / Grayscale Check history does not display the clock time
correctly.
• Perform monitor detection using RadiCS / RadiCS LE. For details, refer to the RadiCS /
RadiCS LE User's Manual.
6.4 Error Code Table
Error Code
****50
****52
****05
****20
****21
[} 27].
• The maximum brightness of the monitor may be lower than the target
brightness.
• Try reducing the target brightness.
• The minimum brightness of the monitor may be higher than the target
minimum brightness.
• Try increasing the target minimum brightness.
• The sensor may not have worked correctly.
• Turn off the main power, wait for a few minutes, turn the power back on, and
then run SelfCalibration / Grayscale Check again.
• The sensor may not have worked correctly.
• Check whether there is any foreign object near the sensor.
• Run SelfCalibration / Grayscale Check again.
Description
Troubleshooting | 6
27