Important Please read PRECAUTIONS, this User’s Manual, and Setup Manual (separate volume) carefully to familiarize yourself with safe and effective usage. • Please refer to the Setup Manual for basic information ranging from connection of the monitor to a PC to using the monitor. • The latest User’s Manual is available for download from our web site: http://www.radiforce.com...
Page 2
EIZO NANAO CORPORATION. EIZO NANAO CORPORATION is under no obligation to hold any submitted material or information confidential unless prior arrangements are made pursuant to EIZO NANAO CORPORATION’s receipt of said information.
Notice for this monitor This product is suited to display medical images of such modalities as MRI, CT, PET, and operating room (RX430-FC only). It does not support the display of mammography images for diagnosis. This product has been adjusted specifically for use in the region to which it was originally shipped. If the product is used outside the region, it may not operate as specified in the specifications.
Cleaning Attention • Never use any solvents or chemicals, such as thinner, benzene, wax, alcohol, and abrasive cleaner, which may damage the cabinet or panel. Note • Optional ScreenCleaner is recommended for cleaning the panel surface. (“ScreenCleaner” is supplied with RX430- FC.) ●...
Chapter 1 Introduction Thank you very much for choosing an EIZO color LCD monitor. 1-1. Features • 29.8 inches • Applicable to the resolution of 4M pixels (2560 × 1600 dots) • Uses high-contrast panel (1100:1). Enables the display of sharp images.
1-2. EIZO LCD Utility Disk An “EIZO LCD Utility Disk” (CD-ROM) is supplied with this product. The following table shows the disk contents and the overview of the software programs. ● Disk contents and software overview The disk includes software programs for adjustment, and User's Manual. Refer to Readme.txt file on the disk for software startup procedures or file access procedures.
1-3. Basic Operations and Functions ● Basic Adjustment menu operations Displaying Adjustment Menu 1. Press . The Adjustment menu appears. Menutitle Menu Adjusting/Setting 1. Choose a menu to adjust/set with , , and then press 2. Choose a menu to adjust/set with , , and then press 3.
● Functions The following table shows all the Adjustment menu’s adjustment and setting menus. Main menu Color Brightness Contrast Temperature Gamma Advanced Settings HybridArea DetectionType HighlightArea ColorReset RadiCSSelfQC SelfCalibration GrayscaleCheck Settings Screen ScreenSize Power PowerSave Manager PresenceSensor PowerIndicator MenuSettings Language MenuPosition Tools Input...
Chapter 2 Adjusting Screens 2-1. Setting the Resolution ● Display resolution For details on compatible resolutions, refer to “Compatible Resolutions” in the Setup Manual. ● Setting resolution When you connect the monitor to the PC and find that the resolution is improper, or when you want to change the resolution, follow the procedure below.
2-2. Adjusting Color ● To select the display mode (CAL Switch mode) This function allows easy selection of a desired mode according to monitor application. Display Modes Mode 1-DICOM AvailabletodisplayinDICOMmode. 2-Custom Selectoneoftheseusermodestodefineadesiredsetofparametersettings. 3-CAL1 Displaysthescreenadjustedbycalibrationsoftware. 4-CAL2 5-CAL3 6-Hybrid-γ Automaticallyidentifiesthedisplayareaforthemonochromeandcolorimagesonthe samescreen,anddisplayseachsetstatus. 7-sRGB SuitableforcolormatchingwithsRGBcompatibleperipherals.
● Adjustable settings in each mode The adjustable settings depend on the mode. (You cannot select any non-adjustable or non-settable function.) √: Adjustable -: Not adjustable Icon Function Brightness Contrast Temperature Gamma Saturation Gain HybridArea DetectionType HighlightArea ColorReset Ifyouadjustthecalibrationinthismodeusingthecalibrationkitexclusivefortothismonitor(seeRadiCS LE/“Accessories”(page38)),only<Brightness>and<ColorReset>areadjustable.
● To adjust contrast <Contrast> The luminance of the screen is adjusted by varying the video signal level. Adjustable range 0 to 100% Procedure 1. Choose <Color> from the Adjustment menu, and press 2. Choose <Contrast> from <Color>, and press The <Contrast>...
● To adjust gamma <Gamma> Gamma can be adjusted. The brightness of the monitor varies depending on the input signal, however, the variation rate is not proportional to the input signal. To keep the balance between the input signal and the brightness of the monitor is called as “Gamma correction”.
● To adjust gain <Gain> The brightness of each color component red, green, and blue is called “Gain”. You can change the hue of “white” by adjusting gain. Adjustable range 0 to 100% Procedure 1. Choose <Color> from the Adjustment menu, and press 2.
● To set the detection type in the monochrome image display area <Detection Type> You can set the detection type in the monochrome image display area. Attention • When the detection type is changed, the display area is highlighted for 5 seconds even if <Highlighted Area> is “Off”. Adjustable range Off, Type-1, Type-2, Type-3 Procedure...
2-3. Screen Size Selection ● To change screen size <Screen Size> The image with the resolution other than the recommended resolution is displayed in full screen automatically. You can change the screen size by using the <Screen Size> function from <Screen>. Setting Full Displaysanimageinfullscreen.Imagesaredistortedinsomecasesbecausethe...
Chapter 3 Setting Monitor 3-1. Enabling/Disabling Mode Selection <Mode Preset> Allows you to select the specified modes only. Use this function when all the display modes are not available or when keeping the display mode unchanged. Procedure 1. Choose <Tools> from the Adjustment menu, and press 2.
3-4. Setting the EIZO Logo Display <Logo> This function allows you to display, or not, the EIZO logo. Procedure 1. Press to turn off the monitor. 2. Holding down, press The Optional Settings menu appears. 3. Choose <Logo> from <Optional Settings>, and press 4.
3-7. Enabling/Disabling DC5V output This function allows you to enable/disable the DC5V output. Procedure 1. Press to turn off the monitor. 2. Holding down, press The Optional Settings menu appears. 3. Choose <DC5V Output> from <Optional Settings>, and press 4. Select ”On” or “Off” with 5.
3-9. Return To Default Setting There are two different Resets: Color Reset for only resetting the color adjustments to the default settings and All Reset for resetting all settings to the defaults. Attention • After resetting, you cannot undo the operation. Note •...
Chapter 4 Controlling Monitor Quality This product has the built-in Integrated Front Sensor, which allows the user to perform periodic calibration and Grayscale Check for the monitor independently. Note • To perform advanced quality control that conforms to medical standards, use the optional monitor quality control software kit “RadiCS UX1”.
4-2. Checking Grayscale <Grayscale Check> This function allows you to execute Grayscale Check and view the latest result. In addition, when the schedule is set using RadiCS, the next grayscale check schedule appears. ● Executing Grayscale Check Procedure 1. Choose <RadiCS SelfQC> from the Adjustment menu, and press 2.
Chapter 5 Power Saving Functions 5-1. Setting Power Saving <Power Save> This function allows you to set the monitor into the power saving mode according to the PC status. When the monitor has shifted to the power saving mode, the images on the screen are not displayed. Attention •...
5-3. Setting the power saving mode when a person leaves the monitor <Presence Sensor> The sensor on the front side of the monitor detects the movement of a person by setting a value from “Level-1” to “Level-4” of the Presence Sensor function. When a person moves away from the monitor, it shifts to the power saving mode automatically and does not display the images on the screen to reduce the power consumption.
Chapter 6 Troubleshooting If a problem still remains after applying the suggested remedies, contact your local dealer. No picture Problem 1. No picture • Powerindicatordoesnotlight. • Powerindicatorislightinggreen. • Powerindicatorislightingorange. • Powerindicatorflashesorangeandgreen. 2. The message below appears. • Thismessageappearswhennosignalis input. Example: Possible cause and remedy •...
Problem • Themessageshowsthattheinputsignalisout ofthespecifiedfrequencyrange.(Suchsignal frequencyisdisplayedinmagenta.) Example: fD: Dotclock fH: Horizontalscanfrequency fV: Verticalscanfrequency Imaging problems Problem 3. The screen is too bright or too dark. 4. Characters are blurred. 5. Afterimages appear. 6. Green/red/blue/white dots or defective dots remain on the screen. 7.
Other problems Problem 8. The adjustment menu does not appear. 9. No mode menu is displayed. 10. (When using Presence Sensor) The images remains to be displayed when you move away from the monitor. The images does not reappear when you come near it.
Chapter 7 Reference 7-1. How to Attach the Optional Arm An optional arm (or an optional stand) can be attached by removing the stand section. (The stand is not supplied with RX430-FC.) Attention • When attaching an arm or stand, follow the instructions of their User’s Manual. •...
7-2. Connecting Multiple PCs The product has multiple connections to PCs and allows you to switch to one of the connections for display. Connection examples ● To switch among input signals <Input> Procedure 1. Choose <Tools> from the Adjustment menu, and press 2.
● To set input signal selection <Input Selection> Setting Auto ThemonitorrecognizestheconnectorthroughwhichPCsignalsareinput. WhenaPCisturnedofforentersthepowersavingmode,themonitorautomatically displaysanothersignal. Manual ThemonitordetectsonlythePC'ssignalscurrentlydisplayingautomatically.Selectan activeinputsignalwith<Input>. Procedure 1. Choose <Tools> from the Adjustment menu, and press 2. Choose <Input Selection> from <Tools>, and press 3. Select “Auto” or “Manual” with 4.
1. A PC equipped with a USB port or another USB hub connected to a USB compatible PC 2. Windows 2000/XP/Vista/7 or Mac OS 9.2.2 and Mac OS X 10.2 or later 3. EIZO USB cable (MD-C93) Attention • This monitor may not work depending on PC, OS or peripheral devices to be used. For USB compatibility of peripheral devices, contact their manufactures.
7-4. Displaying monitor information ● Displaying signal information <Signal Info> This function displays the information about the current input signals displayed. Procedure 1. Choose <Tools> from the Adjustment menu, and press 2. Choose <Signal Info> from <Tools>, and press The Signal Info screen appears. (Example) ●...
7-6. Glossary DDC (Display Data Channel) VESA provides the standardization for the interactive communication of the setting information, etc. between a PC and the monitor. DICOM (Digital Imaging and Communication in Medicine) The DICOM standard was developed by the American College of Radiology and the National Electrical Manufacturer’s Association of the USA.
HDCP (High-bandwidth Digital Content Protection) Digital signal coding system developed to copy-protect the digital contents, such as video, music, etc. This helps to transmit the digital contents safely by coding the digital contents sent via the DVI or HDMI connector on the output side and decoding them on the input side. Any digital contents cannot be reproduced if both of the equipments on the output and input sides are not applicable to HDCP system.
PSP and PS3 are trademarks of Sony Computer Entertainment Inc. EIZO, the EIZO Logo, ColorEdge, DuraVision, FlexScan, FORIS, RadiForce, RadiCS, RadiNET, and ScreenManager are registered trademarks of EIZO NANAO CORPORATION in Japan and other countries. ColorNavigator, EIZO EasyPIX, EcoView NET, EIZO ScreenSlicer, i • Sound, Screen Administrator, UniColor Pro, and Raptor are trademarks of EIZO NANAO CORPORATION.
Medical Standard • It shall be assured that the final system is in compliance to IEC60601-1-1 requirement. • Power supplied equipment can emit electromagnetic waves, that could influence, limit or result in malfunction of the monitor. Install the equipment in a controlled environment, where such effects are avoided.
EMC Information Essential performance of RadiForce series is to display images and operate functions normally. CAUTION The RadiForce series requires special precautions regarding EMC and need to be installed, put into service and used according to the following information. Do not use any cables other than the cables that provided or specifi ed by us. Using other cables may cause the increase of emission or decrease of immunity.
Page 46
Recommended separation distances between portable and mobile RF communications equipment and the RadiForce Series TheRadiForceseriesisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbances arecontrolled.ThecustomerortheuseroftheRadiForceseriescanhelppreventelectromagneticinterferenceby maintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andthe RadiForceseriesasrecommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment. Ratedmaximumoutput poweroftransmitter 150kHzto80MHz d=1.2√ P 0.01 0.12 0.38 For transmitters rated at a maximum output power not listed above, the recommended separation distance “d” in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where “P”...
Need help?
Do you have a question about the RADI FORCE RX430 and is the answer not in the manual?