Analyzer Space; Installation Configuration - Micromeritics ACCUPYC III Pre-Installation Instructions

Gas displacement pycnometer
Hide thumbs Also See for ACCUPYC III:
Table of Contents

Advertisement

2 Pre-installation Instructions
A
S
NALYZER
PACE
An unobstructed lab work space that will accommodate the following specifications is needed:
S
R
PACE
EQUIRED FOR INSTALLATION
Physical
Height
Width
Depth
Weight
I
C
NSTALLATION
Standard installation requires the use of 1/8 in. copper or stainless steel gas supply lines, located
in the instrument accessories kit. A nonstandard installation will be created if another gas supply
line is used or if the gas cylinders cannot be placed within 6 ft (2 m) of the analyzer. There may be
additional costs associated with a nonstandard installation. Please contact the Micromeritics
Service Department at 1-770-662-3636 to discuss a nonstandard installation.
2 - 2
20.5 cm (8.1 in.)
26.5 cm (10.4 in.)
43.0 cm (16.9 in.)
11.5 kg (25.3 lbs.)
ONFIGURATION
AccuPyc III Pre-installation Instructions and Checklist
135-42870-01 (Rev -) — Nov 2023