Philips V60 Developer's Manual page 12

Communications interface for vrpt/snda
Hide thumbs Also See for V60:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Field
Description
58
Unused
59 to 80
Unused
81
Inhalation pressure
setting
82
Inhalation time setting
83 to 88
Unused
89
Alarm silence status
90
Unused
91
Occlusion alarm status
or I-time too long alarm
status
92 to 95
Unused
96
Parameter control setting
97
Unused
98
Stop code
Table 1-2: SNDA record format (continued)
Example
Resolution
◆◆◆◆◆◆◆◆◆
N/A
◆◆◆◆◆◆
◆◆◆◆◆◆
N/A
12.00◆
0.01
0.10◆◆◆
0.01
◆◆◆◆◆◆
N/A
ON◆◆◆◆
N/A
◆◆◆◆◆◆
N/A
NORMAL
N/A
◆◆◆◆◆◆
N/A
I-TIME◆◆◆
N/A
◆◆◆◆◆◆
N/A
0x03
N/A
Range
Units
Comments
N/A
N/A
Always output as
"◆◆◆◆◆◆◆◆◆
◆◆◆◆◆◆"
N/A
N/A
Always output as
"◆◆◆◆◆◆"
4.00 to
cmH
O
IPAP in PCV mode
2
40.00
IPAP in S/T mode
CPAP in CPAP mode
"◆◆◆◆◆◆" in other
modes
0.10 to 3.00
sec
I-Time in PCV mode
"◆◆◆◆◆◆" in other
modes
N/A
N/A
Always output as
"◆◆◆◆◆◆"
ON◆◆◆◆
N/A
OFF◆◆◆
N/A
N/A
Always output as
"◆◆◆◆◆◆"
NORMAL
N/A
Report highest
urgency of these
RESET◆
alarms: Patient
ALARM◆
Circuit Occluded and
Patient Disconnect.
In HFT, these alarms:
"Patient Circuit
Occluded" and
"Cannot Reach Target
Flow."
N/A
N/A
Always output as
"◆◆◆◆◆◆"
N/A
N/A
"I-TIME◆◆◆" in
Ventilation modes.
"◆◆◆◆◆◆◆◆◆" in
HFT.
N/A
N/A
Always output as
"◆◆◆◆◆◆"
N/A
N/A
ASCII End
Transmission
character (ETX)
11