Download Print this page

Philips V60 Developer's Manual page 4

Communications interface for vrpt/snda
Hide thumbs Also See for V60:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Field
Description
57
NPPV IPAP setting
(or CPAP setting).
58
NPPV inspiratory time
setting
59
NPPV rise time
60
NPPV I-trigger type
61
Unused
62
NPPV E-cycle type
63
Unused
64
NPPV oxygen
concentration setting
65
Unused
66
NPPV low inspiratory
pressure alarm limit
setting
67
Unused
68
NPPV low tidal volume
alarm limit setting
69
NPPV high respiratory
rate alarm limit setting
70
NPPV low minute volume
alarm limit setting
71 to 72
Unused
Table 1-1: VRPT record format (continued)
Example
Resolution
5◆◆◆◆◆
1
1.00◆◆
0.05
0.1◆◆◆
0.1
AUTO◆◆
N/A
◆◆◆◆◆◆
N/A
AUTO◆◆
N/A
◆◆◆◆◆◆
N/A
21◆◆◆◆
1
N/A
◆◆◆◆◆◆
3◆◆◆◆◆
1
◆◆◆◆◆◆
N/A
0◆◆◆◆◆
5
60◆◆◆◆
1
1.00◆◆
0.01 for
0.00 to
9.99
0.1 for 10.0
to 99.0
◆◆◆◆◆◆
N/A
Range
Units
Comments
4 to 40
cmH
O
IPAP setting in S/T
2
and PCV
CPAP setting in CPAP
mode
In CPAP mode:
EPAP = IPAP = CPAP.
"◆◆◆◆◆◆" in other
modes
0.30 to 3.00
sec
"◆◆◆◆◆◆" in HFT
0.1 to 0.5
N/A
"◆◆◆◆◆◆" in HFT
(Display range 1-5).
N/A
N/A
"◆◆◆◆◆◆" in HFT
N/A
N/A
Always output as
"◆◆◆◆◆◆"
N/A
N/A
"◆◆◆◆◆◆" in HFT
N/A
N/A
Always output as
"◆◆◆◆◆◆"
21 to 100
%
N/A
N/A
Always output as
"◆◆◆◆◆◆"
0 to 40
cmH
O
Off = 0
2
"◆◆◆◆◆◆" in HFT
N/A
N/A
Always output as
"◆◆◆◆◆◆"
0 to 1500
mL
Off = 0
"◆◆◆◆◆◆" in HFT
5 to 90
BPM
"◆◆◆◆◆◆" in HFT
0 to 99
L/min
"◆◆◆◆◆◆" in HFT
N/A
N/A
Always output as
"◆◆◆◆◆◆"
3