SITRANS FM100
Out 1 configured as switching output
pushpull and assigned to flow
Font white: MV within MR
Font yellow: 100% MR <=MV <=OverFlow
Font red: MV OverFlow
Legend:
MV - measured value
MR - measuring range
Out 1 configured as pulse output and
assigned to partial totalizer
9.2.1
The number of digits displayed on the volume counter display (partial and total
volumetric counters) is limited to max. 8 digits. The partial and total volumetric
meters therefore have a smaller font size than the flow and temperature display.
If the 8-digit display range of the meter is exceeded, this is indicated by the
display of 8 minus characters (--------). In this case, the meter reading can no
longer be read. The user now has the option of bringing the counter reading back
into the display area by changing the volume counter unit.
page 14
Measurement Mode Display Layout 'Dual'
Display area of the flow meters
Out 2 configured as analogue output 4-20 mA
and assigned to temperature
Measuring variables with sign for direction
Unit for measuring variable
Out 2 configured as analogue output 4-20 mA
and assigned to temperature
A5E48762204-AB