set gnss mgadbd [y | n]
set gnss offline [y | n]
5.11 Power commands
Command
poweroff
hibernate <minutes>
set wakeonpower [y/n]
set powerofftimeout <minutes>
set batwarning <voltage>
set batshutdn <voltage>
set batvnoboot <voltage>
AML Oceanographic Ltd.
setting to determine its location if in manual location mode.
Enable/disable GPS database dump to enable GPS warm
start. If enabled, the GPS module will store almanac data
that will improve lock times after a recent fix has been lost.
Enabling this feature generally improves GPS lock times.
Enable/disable GPS offline assist. Requires connecting to
Sailfish with an active internet connection at least once
every 5 days. Offline data will help improve lock times on
power up.
Description
Turn the instrument off indefinitely, until the power dial
switch is toggled or external power cycled.
Turn the instrument off for a set amount of minutes.
Once enabled, the instrument will stay on until a Power Off
command is issued or poweroff timeout has occurred.
Note: This is a secure command.
Set the instrument's idle timeout in minutes. If there is no
user interaction and the instrument is idling, i.e. not
streaming data or logging, the instrument will power down
after the set timeout.
Setting the timeout to 0 disables this feature.
Set the threshold for the low battery warning to be triggered.
If triggered the indicator led will turn red to notify users to
charge the instrument.
Note: This is a secure command.
Set the threshold for the low battery shutdown to be
triggered. If the instrument's battery voltage drops below this
threshold the instrument will shut down.
It is not recommended to set this lower than 6.2 V.
Note: This is a secure command.
On power up if the battery voltage is less than the
batvnoboot value, the instrument will not boot up.
31 / 43
Need help?
Do you have a question about the AML-3 and is the answer not in the manual?