AML Oceanographic AML-3 User Manual page 34

Multiparameter logger instrument
Table of Contents

Advertisement

Sample output in this format is shown below.
>MONITOR
>msg138{mux[meta=time,1590605500.55,s],port1[data=Cond,0.000000,mS/cm][rawi=ADC,563,none][dat
a=TempCT,23.881313,C][rawi=ADC,428710,none],port2[data=Pressure,0.071390,dbar][rawi=ADC,84447
0,2sComp],port3[data=SV,0.000000,m/s][rawf=NSV,0.000000,samples],derive[data=Depth,0.070998,m]}
In this format every measurement has, at minimum, an explicit unit, parameter, and port
association. A driver written to interpret this format is able to understand output from every
X2Change generation instrument regardless of specific sensor or derived parameter
configuration. Users are strongly encouraged to use drivers that rely on this output
format.
Parameter name
SV
TempSVT
Cond
TempCT
Pressure
Turbidity
Dissolved Oxygen
DOM
Chloro-blue
Chloro-red
pH
Fluorescein
Oil-crude
Oil-fine
Brighteners
Phycocyanin
Phycoerythrin
PTSA
AML Oceanographic Ltd.
Description
Sound velocity
Temperature from an SVT.X2 sensor
Conductivity
Temperature from a CT.X2 sensor.
Pressure
Turbidity
Dissolved Oxygen
CDOM/FDOM
Chlorophyll, blue excitation
Chlorophyll, red excitation
pH
Fluorescein dye
Crude oil
Fine oil
Optical brighteners
Phycocyanin (freshwater cyanobacteria)
Phycoerythrin (marine cyanobacteria)
PTSA
34 / 43