5 Appendix A - Full command list
All of the commonly utilized settings are accessible via GUI using Sailfish. However, Sailfish
does not provide GUI access to the entire command set; advanced functionality beyond
common settings exists. This advanced functionality is described here.
These commands are applied to the instrument via command line. This is achieved by using the
Advanced Commands section of the bottom of the Instrument Settings page in Sailfish.
5.1 Sampling commands
Command
set sample <rate> [/second | /minute, /hour |
seconds | minutes | hours | max]
set scan mode [burst | single]
set burst interval <interval>
set burst samples <samples>
set burst monitor [samples | average]
display burst
set trigger [time | sound | pressure]
AML Oceanographic Ltd.
Description
Set the rate at which the instrument logs and
monitors data from the sensors. Maximum
sample rate is 20 Hz.
Switch between single and burst mode. When
the instrument is in burst mode it will sample
in bursts at a set interval. Note that the burst
duration must be less than or equal to the
burst interval.
Set the burst interval in seconds.
Minimum: 5
Maximum: 14400
Set the number of samples taken per burst.
Samples will be taken at the set sample rate.
Minimum: 2
Maximum: 1200
Set whether the monitor output will display all
the burst samples or an average of the
samples.
Note: averaging applies only to monitor data.
Log files will record individual samples.
Display the current burst mode settings.
Set the sampling trigger to sample by time,
sound velocity increment or pressure
increment.
Sampling by sound or pressure increment is
not supported in burst mode.
24 / 43
Need help?
Do you have a question about the AML-3 and is the answer not in the manual?