Pickering 40-518A User Manual page 10

Pxi 2 a 2-pole matrix module
Table of Contents

Advertisement

SECTION 1 - TECHNICAL SPECIFICATION
Specifications
Spare Relay
A Spare Relay is built onto the daughter board to allow
easy maintenance with minimum downtime.
Switching Specification
Switch Type:
Contact Type:
Max Switch Voltage:
Max Power:
Max Switch Current:
Max Continuous Carry Current:
Max Pulsed Carry Current
Example
(for a single switch path):
Initial On Path Resistance:
Off Path Resistance:
Minimum Voltage
Differential Thermal Offset:
Operate Time:
Expected Life (Operations)
Very low power load:
Low power load:
Medium power load:
Full power load:
* For full voltage rating, signal sources to be switched
must be fully isolated from mains supply and safety earth.
Bandwidth
Typical Bandwidth:
Power Requirements
+3.3 V
+5 V
100 mA
400 mA
(Typical)
(Typical)
Page 1.2
Electro-mechanical
Palladium-Ruthenium,
Gold plated, bifurcated
300 VDC/250 VAC*
62.5 VA, 60 W from 30 V to
220 VDC, 30 W to 300 VDC
(resistive load)
2 A
2 A
6 A for 100 ms
(up to 10 % duty cycle)
<500 mΩ
>10
9
100 µV
<10 µV
<3 ms
>1x10
8
>1.5x10
(0.1 A 20V DC)
7
>5x10
(1 A 30V DC)
6
>1x10
(2 A 30V DC)
5
>1x10
(0.1 A 300 VDC)
5
>10 MHz
+12 V
-12 V
50 mA
0
(Typical)
2 A 2-POLE MATRIX MODULE 40-518A
Mechanical Characteristics
Single slot 3U PXI (CompactPCI card).
3D models for all versions in a variety of popular file
formats are available on request.
Connectors
PXI bus via 32-bit P1/J1 backplane connector.
Signals via front panel 78-pin male D-type connector, for
pin outs please refer to the operating manual.
Note: The pinout for the 40-518A differs from the 40-518,
see the user manual for details.
Operating/Storage Conditions
Operating Conditions
Operating Temperature:
Humidity:
Altitude:
Storage and Transport Conditions
Storage Temperature:
Humidity:
Altitude:
PXI & CompactPCI Compliance
The module is compliant with the PXI Specification 2.2.
Local Bus, Trigger Bus & Star Trigger are not implemented.
Uses a 33 MHz 32-bit backplane interface.
Safety & CE Compliance
All modules are fully CE compliant and meet
applicable EU directives: Low-voltage safety EN61010-
1:2010, EMC Immunity EN61326-1:2013, Emissions
EN55011:2009+A1:2010.
pickering
2 A 2-Pole Matrix, 40-518A
0° C to +55° C
Up to 90 % non-condensing
5000 m
-20 °C to +75 °C
Up to 90 % non-condensing
15000 m
pickeringtest.com

Advertisement

Table of Contents