Using The Centrifuge; Initial Start-Up; Switching The Centrifuge On; Opening And Closing The Lid - Sigma 4-16KS Operating Manual

Refrigerated centrifuge
Hide thumbs Also See for 4-16KS:
Table of Contents

Advertisement

Refrigerated Centrifuge Sigma 4-16KS

6 Using the centrifuge

Pos : 105 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/060 Betrieb====================== =================== @ 27\mod_1405319198895_68.docx @ 192423 @ 1 @ 1
6
Using the centrifuge
Pos : 106 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/060- 0010 Erste Inbetriebnahme @ 27\m od_1405319200130_68.docx @ 192437 @ 2 @ 1
6.1

Initial start-up

DANGER
Pos : 107 /010 Univ ers almodul e/
Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 108 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/060- 0020 Ei nsc halten (auß er 6- 16HS) @ 27\mod_1405319201340_68.docx @ 192451 @ 2 @ 1
6.2

Switching the centrifuge on

Pos : 109 /010 Univ ers almodul e/
Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 110 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/060- 0020-0010 Öffnen und Sc hließ en des Dec kels - 1 Sc hloss @ 27\m od_1405319202508_68.docx @ 192465 @ 3 @ 1
6.2.1

Opening and closing the lid

WARNING
Pos : 111 /010 Univ ers almodul e/
Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 112 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/060- 0020-0019 Ei ns etz en v on Rotor en und Zubehör ( auß er 8K)-------------------------------- @ 27\m od_1405319204492_68.docx @ 192493 @ 3 @ 1
6.2.2

Installation of rotors and accessories

Pos : 113 /100 Sigma/110 BA Z entrifugen Sigma (Proj ekte)/4-16KS_4- 16KHS/060 Betri eb/060-0020-0020- 0010 Ei nsetzen ei nes Rotors 4- 16KS_KHS @ 29\mod_1405326945297_68.docx @ 198765 @ 4 @ 1
6.2.2.1

Installation of a rotor

WARNING
34 / 99
Before the initial start-up, please ensure that your centrifuge is
properly set up and installed (see chapter 5 - "Set-up and
connection").
Press the mains power switch.
The display then illuminates. The centrifuge is ready for operation.
The lid can be opened if the centrifuge is at a standstill and if the lid key is
illuminated.
Press the lid key in order to open the lid.
The centrifuge cannot be started if the lid is opened.
To close, press with both hands slightly on the lid until the lock is
audibly locked.
Do not place your fingers between the lid and the housing when closing
the lid. Risk of crushing!
Open the centrifuge lid by pressing the lid key.
Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
Lower the rotor with its central bore straight down onto the motor shaft.
Tighten the rotor tie-down screw clockwise with the supplied rotor
wrench with 10 Nm (Option: torque wrench 10 Nm, part no. 17060). In
doing so, hold the rotor at its outer rim.
Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
Once a day or after 20 cycles, the rotor tie-down screw must be loosened
by some turns, and the rotor must be lifted and fastened again. This
ensures a proper connection between the rotor and the motor shaft.
Version 07/2012, Rev. 1.26 of 24/10/2023 • sb
Original operating manual, part no. 0701702