Clear and Sanitize Procedures

Memory Devices

Table 1: Volatile Memory Devices for the DPO2000 and MSO2000 Series
Type and minimum size
CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84
CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84
CMOS, SDRAM, 64MX16,
1G DDR2, 1.8V, 333MHZ;
MT47H64M16BT-3, FBGA84
CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84
Table 2: Nonvolatile Memory Devices for the DPO2000 and MSO2000 Series
Type and minimum size
CMOS, FLASH; 16M X 8 OR
8M X 16, 128MBIT, 75NS, 3V;
PC28F128J3D-75, BGA64
DPO2000 and MSO2000 Series Declassification and Security Instructions
The following tables list the volatile and nonvolatile memory devices in the
standard instrument and listed options.
Function
User modifiable
DDR memory: contains
Only by the
the analog channel
processor system
waveform acquisition
data during operation
DDR memory: contains
Only by the
the digital channel
processor system
waveform acquisition
data during operation
DDR memory: contains
Only by the
the instrument code and
processor system
data during operation
DDR memory: contains
Only by the
the instrument code and
processor system
data during operation
Function
User modifiable
FLASH: contains the
N/A
instrument firmware,
calibration constants,
reference waveforms,
and instrument setups
Data input
method
Location
Acquisition board
N/A
N/A
Acquisition board
(MSO2000 Series only)
N/A
Acquisition board
Acquisition board
N/A
(4-channel products
only)
Data input
method
Location
N/A
Acquisition board
To clear or sanitize
Remove power from the
instrument for at least
20 seconds
Remove power from the
instrument for at least
20 seconds
Remove power from the
instrument for at least
20 seconds
Remove power from the
instrument for at least
20 seconds
To clear or sanitize
To clear: Not applicable
for calibration constants.
Use the TekSecure
feature to erase
reference waveforms
and instrument setups.
Use the procedure to
disable the LAN Ethernet
and to clear information,
such as IP addresses.
(See page 6, Disable the
LAN Port and Clear LAN
Ethernet Settings.)
To sanitize: Remove
the acquisition board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Return the product
to Tektronix and a
new acquisition board
will be installed and
the instrument will be
repaired and adjusted as
necessary.