Disable The Lan Port And Clear Lan Ethernet Settings - Tektronix DPO2014 Instructions Manual

Oscilloscopes
Hide thumbs Also See for DPO2014:
Table of Contents

Advertisement

Built-In Security Features

Disable the LAN Port and Clear LAN Ethernet Settings

6
To reset the instrument RAM:
1. Power off the instrument for at least 20 seconds.
2. Power on the instrument.
To disable the LAN port and clear LAN Ethernet settings, such as IP addresses,
follow these steps:
CAUTION.
If you need to restore network connectivity at a later date, make sure to
write down all setting values before clearing them.
1. Remove the network cable from the LAN port on the rear of the instrument.
2. Push the front-panel Utility button.
3. Push the Utility Page lower-bezel button and use the Multipurpose a knob
to select I/O.
4. Push the Ethernet Network Settings lower-bezel button.
5. Push the Change Instrument Settings side-bezel button.
6. Push the ↑ or ↓ arrow side-bezel button to position the cursor on the User
Password information.
7. Push the Clear lower-bezel button.
8. Push the ↓ arrow side-bezel button to position the cursor on the Instrument
IP Address information.
9. Push the Clear lower-bezel button.
10. Push the ↓ arrow side-bezel button to position the cursor on the Domain
Name information.
11. Push the Clear lower-bezel button.
12. Push the ↓ arrow side-bezel button to position the cursor on the DNS IP
Address information.
13. Push the Clear lower-bezel button.
DPO2000 and MSO2000 Series Declassification and Security Instructions