Data Export Devices

Table 3: Data Export Devices
Type and minimum size
USB host port
USB device port
DPO2000 and MSO2000 Series Declassification and Security Instructions
The following table lists the data export devices.
Function
User modifiable
Yes
Supports the removable
USB flash drive; user
storage of reference
waveforms, screen
images, and instrument
setups
Supports remote control
Yes
and data transfer to a
PC
Clear and Sanitize Procedures
Data input
method
Location
Save
USB host port on the
front of the instrument
Remote
USB device port on the
control via
rear of the instrument
USBTMC
To disable
Files can be deleted
or overwritten on the
instrument or a PC; the
USB flash drive can be
removed and destroyed.
The USB host port
cannot be disabled.
Use the procedure to
disable the USB device
port. (See page 7,
Disable the USB Device
Port.)