Monitoring Both Phases Of A Status Transition; Inhibit/Fault Indicator; Remote Inhibit (Ri); Discrete Fault Indicator (Dfi) - Keysight Technologies 66332A Programming Manual

Table of Contents

Advertisement

3 - Programming the DC Source
STATus:QUEStionable:PTR 19;ENABle 19
Step 3
Program the Service Request Enable register to allow both the Operation and the
Questionable summary bits from the Status Byte register to generate RQS. Use:
*SRE 136
Step 4
When you service the request, read the event registers to determine which Operation
Status and Questionable Status Event register bits are set, and clear the registers for
the next event. Use:
STATus:OPERation:EVENt;QUEStionable:EVENt?

Monitoring Both Phases of a Status Transition

You can monitor a status signal for both its positive and negative transitions. For example, to generate
RQS when the dc source either enters the CC+ (constant current) condition or leaves that condition,
program the Operational Status PTR/NTR filter as follows:
STATus:OPERational:PTR 1024;NTR 1024
STATus:OPERational:ENABle 1024;*SRE 128
The PTR filter will cause the OPERational summary bit to set RQS when CC+ occurs. When the
controller subsequently reads the event register with STATus:OPERational:EVEN?, the register is cleared.
When CC+ subsequently goes false, the NTR filter causes the OPERational summary bit to again set
RQS.

Inhibit/Fault Indicator

The remote inhibit(INH) and discrete fault(FLT) indicators are implemented through the respective INH
and FLT connections on the rear panel. Refer to Table 1-2 for the electrical parameters.

Remote Inhibit (RI)

Remote inhibit is an external, chassis-referenced logic signal routed through the rear panel INH
connection, which allows an external device to signal a fault. To select an operating modes for the remote
inhibit signal, use:
OUTPut:RI:MODE LATChing | LIVE | OFF

Discrete Fault Indicator (DFI)

The discrete fault indicator is an open-collector logic signal connected to the rear panel FLT connection,
that can be used to signal external devices when a fault condition is detected. To select the internal fault
source that drives this signal, use:
OUTPut:DFI:SOURce QUEStionable | OPERation | ESB | RQS | OFF
To enable or disable the DFI output, use:
OUTPut:DFI:STATe ON | OFF
36
(8 + 128 = 136)
(1 + 2 + 16 = 19)

Advertisement

Table of Contents
loading

This manual is also suitable for:

66312a6631b6632b6633b6634b6611c ... Show all

Table of Contents