Calibration Commands; Calibrate:current; Calibrate:current:negative; Calibrate:current:measure:lowrange - Keysight Technologies 66332A Programming Manual

Table of Contents

Advertisement

4 - Language Dictionary

Calibration Commands

Calibration commands let you:
 Enable and disable the calibration mode
 Change the calibration password
 Calibrate the current and voltage programming and measurement, and store new calibration
constants in nonvolatile memory.
NOTE:
If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error.

CALibrate:CURRent

This command initiates the calibration of the positive dc output current as well as the high-range current
measurement circuit.
Command Syntax
Parameters
Examples
Related Commands

CALibrate:CURRent:NEGative

This command initiates the calibration of the negative dc output current.
Command Syntax
Parameters
Examples
Related Commands

CALibrate:CURRent:MEASure:LOWRange

This command initiates the calibration of the low-range current measurement circuit.
Command Syntax
Parameters
Examples
Related Commands

CALibrate:CURRent:MEASure:AC

Keysight 66312A, 66332A Only
This command initiates the calibration of the high bandwidth (ac) measurement circuit.
Command Syntax
Parameters
Examples
44
CALibrate:CURRent[:SOURce][:DC][:POSitive]
None
CAL:CURRCAL:CURR:SOUR:DC:POS
CAL:CURR:NEG
CALibrate:CURRent[:SOURce][:DC]:NEGative
None
CAL:CURR:NEGCAL:CURR:SOUR:DC:NEG
CAL:CURR
CALibrate:CURRent:MEASure[:DC]:LOWRange
None
CAL:CURR:MEAS
CAL:CURR
CALibrate:CURRent:MEASure:AC
None
CAL:CURR:MEAS:AC

Advertisement

Table of Contents
loading

This manual is also suitable for:

66312a6631b6632b6633b6634b6611c ... Show all

Table of Contents