Vega VEGAMIP R62 Operating Instructions Manual page 42

Table of Contents

Advertisement

9 Supplement
Output variable
Output
Load current
Voltage loss
Switching voltage
Blocking current
Modes (switchable)
Measurement accuracy
Hysteresis
Switching delay
Min. distance
Ambient conditions
Ambient, storage and transport tempera-
ture
Process conditions
Measured variable
Process pressure
4)
Ʋ VEGAMIP 62, threaded version, inter-
nal horn antenna with PTFE cover
Ʋ VEGAMIP 62, plastic encapsulated
antenna with PP cover
Ʋ VEGAMIP 62, horn antenna (316L)
Ʋ VEGAMIP 62, encapsulated horn
antenna with PTFE cover
Ʋ VEGAMIP 62 with mounting adapter
"Abrasion protection"
Ʋ VEGAMIP 62 with mounting adapter
"High temperature" 150 mm
Ʋ VEGAMIP 62 with mounting adapter
"High temperature" 300 mm
Process temperature (thread or flange
temperature)
Ʋ VEGAMIP 62, threaded version, inter-
nal horn antenna with PTFE cover
Ʋ VEGAMIP 62, plastic encapsulated
antenna with PP cover
Ʋ VEGAMIP 62, horn antenna/316L -
seal: FKM (SHS FDM 70C3 GLT)
Ʋ VEGAMIP 62, horn antenna/316L -
seal: FFKM (Kalrez 6375)
Not overload resistant and not short-circuit proof
3)
Note max. pressure of the process fitting.
4)
42
Floating transistor output
< 400 mA
< 0.5 V
< 55 V DC
< 10 µA
min./max.
approximately 1 dB
adjustable from 0.1 ... 20 s (on/off)
100 mm (3.94 in)
-40 ... +80 °C (-40 ... +176 °F)
Limit level of bulk solids and liquids
-1 ... 4 bar/-100 ... 400 kPa (-14.5 ... 58 psig)
-1 ... 2 bar/-100 ... 200 kPa (-14.5 ... 29 psig)
-1 ... 40 bar/-100 ... 4000 kPa (-14.5 ... 580 psig)
-1 ... 16 bar/-100 ... 1600 kPa (-14.5 ... 232 psig)
-1 ... 20 bar/-100 ... 2000 kPa (-14.5 ... 290 psig)
unpressurized (IP67)
unpressurized (IP67)
-40 ... +80 °C (-40 ... +176 °F)
-40 ... +80 °C (-40 ... +176 °F)
-40 ... +130 °C (-40 ... +266 °F)
-20 ... +130 °C (-4 ... +266 °F)
3)
VEGAMIP R62 • Receiving unit