Vega VEGAMIP R62 Operating Instructions Manual page 16

Table of Contents

Advertisement

4 Mounting
Threaded version
16
Fig. 9: Installation in pipelines
x
Min. distance 100 mm (3.94 in)
Mount the threaded version of VEGAMIP 62 in the following way:
Fig. 10: VEGAMIP 62, threaded version G1½
1
Polarisation marking
2
Instrument hexagon
3
Counter nut
4
Process seal
1. Insert the supplied process seal (4) before screwing in the instru-
ment
2. To screw VEGAMIP 62 in, use the instrument hexagon (2) below
the housing.
Screw the instrument in and tighten it lightly with a flat spanner
3. Turn the sensor back (max. 180°) to reach the required orientation
of the polarisation marking (1)
4. Hold the instrument in this position and tighten with the counter
nut (3)
Caution:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
Note:
In the case of VEGAMIP 62 with NPT thread, the instrument seals in
the thread itself. Hence, no counter nut is necessary for these ver-
sions.
x
1
2
3
4
VEGAMIP R62 • Receiving unit

Advertisement

Table of Contents
loading

Table of Contents