Output Characteristic - Keysight Technologies 66312A User Manual

Dynamic measurement dc source system, dc power supplies
Hide thumbs Also See for 66312A:
Table of Contents

Advertisement

2 - General Information

Output Characteristic

The dc source's output characteristic is shown in the following figure. The output of the dc source may be
adjusted to any value within the boundaries shown.
-
The dc source can operate in either constant voltage (CV) or constant current (CC) over the rated output
voltage and current. Although the dc source can operate in either mode, it is designed as a constant
voltage source. This means that the unit turns on in constant voltage mode with the output voltage rising
to its Vset value. There is no command for constant current operation. The only way to turn the unit on in
constant current mode is by placing a short across the output and then enabling or turning the output on.
Note that the dc source cannot be programmed to operate in a specific mode. After initial turn-on, the
operating mode of the unit will be determined by the voltage setting, the current setting, and the load
resistance. In figure 2-1, operating point 1 is defined by the load line traversing the positive operating
quadrant in the constant voltage region. Operating point 2 is defined by the load line traversing the
positive operating quadrant in the constant current region.
Figure 2-1 also shows a single range  two quadrant capability. This means that the dc source is capable
of sourcing as well as sinking current over the output voltage range from zero volts to the rated
maximum. The negative current sinking capability of the dc source is not programmable, and is fixed at a
maximum of approximately 60% of Imax.
NOTE:
If you attempt to operate the dc source beyond its output ratings, the output of the unit
will become unregulated. This is indicated by the UNR annunciator on the front panel.
The output may also become unregulated if the ac input voltage drops below the
minimum rating specified in Appendix A.
Appendix A documents the dc source's specifications and supplemental characteristics.
20
Output
Voltage
Vmax
VSET
-Isink
0
(60% Imax)
Figure 2-1. Dc Source Output Characteristic
CV operating line
1
2
CC operating line
+
ISET
Imax
Output
Current

Advertisement

Table of Contents
loading

This manual is also suitable for:

6611c6612c6613c6614c

Table of Contents