Front Panel Menus - At A Glance - Keysight Technologies 66312A User Manual

Dynamic measurement dc source system, dc power supplies
Hide thumbs Also See for 66312A:
Table of Contents

Advertisement

Front Panel Menus - At a Glance

Address
ADDRESS 7
INTF GPIB
BAUDRATE 300
PARITY NONE
FLOW NONE
LANG SCPI
*RCL 0
Recall
*SAV 0
Shift
Save
ERROR 0
Shift
Error
12.000V 0.204A
Meter
12.500V MAX
1.000V MIN
12.330V HIGH
0.080V LOW
12.000V RMS
0.350A MAX
0.050A MIN
0.400A HIGH
0.012A LOW
0.210A RMS
VOLT 20.000
Voltage
CURR 2.000
Current
OC -- -- -- --
Protect
*RST
Output
PON:STATE RST
PROT:DLY 0.08
RI LATCHING
DFI OFF
DFI:SOUR OFF
PORT RIDFI
DIGIO 7
RELAY ON
RELAY NORM
VOLT:PROT 22
Shift
OV
CURR:RANG HIGH
Shift
Input
CURR:DET ACDC
CAL ON
Shift
Cal
Use
and
to select menu parameters.
1
Not available on Keysight 6611C – 6614C or in Compatibility mode.
Sets the GPIB Address
Selects an interface (GPIB or RS232)
Selects baud rate (300, 600, 1200, 2400, 4800, 9600)
Selects message parity (NONE, EVEN, ODD, MARK, SPACE)
Selects flow control (XON-XOFF, RTS-CTS, DTR-DSR, NONE)
Selects language (SCPI or COMP)
Recalls instrument state
Saves present instrument state
Displays errors in SCPI error queue
Measures output voltage and current
Measures peak output voltage
Measures minimum output voltage
Measures the high level of a voltage pulse waveform
Measures the low level of a voltage pulse waveform
Measures rms voltage
1
Measures peak output current
Measures minimum output current
Measures the high level of a current pulse waveform
Measures the low level of a current pulse waveform
Measures rms current
1
Sets the output voltage
Sets the output current
Protection status (example shows overcurrent tripped)
Places the dc source in the factory-default state
Select the power-on state command (RST or RCL0)
Sets the output protection delay in seconds
Sets the remote inhibit mode (LATCHING, LIVE, or OFF)
Sets the discrete fault indicator state (ON or OFF)
Selects the DFI source (QUES, OPER, ESB, RQS, or OFF)
Sets the output port functions (RIDFI or DIGIO)
Sets and reads the I/O port value (0 through 7)
Sets the output relay state (ON or OFF)
Sets the output relay polarity (NORM or REV)
Sets overvoltage protection level
Sets current range (HIGH, LOW, or AUTO)
Sets current measurement detector (ACDC or DC)
Accesses calibration menu (See User's Guide).
Use
Meter
to exit any menu and return to metering mode.
Quick Reference - 1
1
1
1
1
1
1
1
1
2
2
1
2
Not available on Keysight 66312A.
13