Maintenance

Exchange the Outlet Valve

Exchange the Outlet Valve

When
if leaking internally
Tools required
p/n
8710-1924
8710-0510
5067-5688
Parts required
#
1
OR
1
Preparations
Switch off the pump at the main power switch
1
Using a ¼ inch wrench disconnect the absorber
capillary from the outlet valve.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Description
A
Open-end wrench 14 mm
A
Open-end wrench 1/4 — 5/16 inch
A
Torque wrench 1 – 25 Nm with 14 mm wrench
p/n
Description
A
G1312-60067
Outlet valve (standard)
A
G1312-60167
Outlet valve (type N/SFC)
2
Unscrew the valve with the 14 mm wrench and remove
it from the pump body.
NOTE
Do not disassemble the outlet valve, as this can damage the
valve.