Motor-Drive Power

Error ID: 2041, 2042
Motor-Drive Power: Pump channel A
B: Motor-Drive Power: Pump channel B
The current drawn by the pump motor exceeded the maximum limit.
Blockages in the flow path are usually detected by the pressure sensor, which
result in the pump switching off when the upper pressure limit is exceeded. If a
blockage occurs before the pressure sensor, the pressure increase cannot be
detected and the module will continue to pump. As pressure increases, the pump
drive draws more current. When the current reaches the maximum limit, the
module is switched off, and the error message is generated.
Probable cause
1
2
3
4
5
6
7
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Flow path blockage in front of the pressure
sensor.
Blocked (passive or active) inlet valve.
Blocked outlet valve.
High friction (partial mechanical blockage) in
the pump drive assembly.
Defective pump drive assembly.
Defective mainboard.
Restriction capillary blocked at pre-mixing
union.
Suggested actions
Ensure the capillaries and frits between the
pump head and pressure sensor inlet are free
from blockage.
Exchange the (passive or active) inlet valve.
Exchange the outlet valve.
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Exchange restriction capillary.