Replace the Outlet Valve

When
If Outlet valve is defective.
Parts required
#
1
OR
1
1
Remove the cap from the outlet valve.
3
Remove the heat exchanger capillary by pushing the
connector up and pulling it out of the valve.
N O T E
A gold seal between outlet valve and heat exchanger
capillary is used for a tight connection.
1290 Infinity II High Speed Pump User Manual
p/n
Description
5067-6642
Check valves, ATP head
G7161-60052
Check valves, Prep head
2
Counter the outlet valve while opening the lock screw of
the heat exchanger capillary.
4
Unscrew the outlet valve with a 14 mm wrench (1) and
remove it (2).
Maintenance

Replace the Outlet Valve