7
Maintenance
Replace the Outlet Valve
Replace the Outlet Valve
When
If Outlet valve is defective.
Parts required
#
1
OR
1
1
Lift up solvent filters in solvent reservoirs to avoid
leakages.
2
Open the purge valve.
3
Power off the system.
4
Open the doors.
6
Unscrew the outlet valve with a 17 mm wrench (1.) and
remove it (2.).
Preparative Binary Pumps User Manual
p/n
Description
5067-6642
Check valves, 50 mL head
G7161-60052
Check valves, 200 mL head
2.
1.
5
Remove all tubings and capillaries from the pump head
assembly (seal wash tubes, IN and OUT capillaries).
7
Insert the outlet valve into the pump head (1.). Close the
outlet valve (2.).
1.
2.
110