8
Maintenance
Exchange the Outlet Valve
Exchange the Outlet Valve
For bio-inert modules use bio-inert parts only!
When
If internally leaking
Tools required
p/n
8710-0510
8710-1924
5067-5688
Parts required
p/n
G1312-60067
G5611-60067
Preparations
• Switch off the pump at the main power switch.
• Open the doors of the module.
144
Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
Description
Open-end wrench 1/4 — 5/16 inch
Open-end wrench 14 mm
Torque wrench 1 – 25 Nm with 14 mm wrench
Description
Outlet valve
Outlet Valve