Keysight M8199B Getting Started Manual page 38

Arbitrary waveform generators
Hide thumbs Also See for M8199B:
Table of Contents

Advertisement

2
Basic Setup for M8100A
M8199A
M8008A
38
Case 3
Connecting M8008A clock module to two M8199A AWG modules
In this case, a M8008A clock module, two M8199A AWG modules, a
M9505A AXIe chassis and the following cables are required:
3.5 mm, 50 cm M8199A-811 customer orderable sync cable
(M8199-61620)
1.85 mm, 45 cm M8199A-810 replacement channel clock cable
(M8199-61624)
Make the connections as described below:
1
Connect the M8008A Clock Generator Sync Out port with both the
M8199A AWG Sync In ports using the sync cable (blue cable).
2
Connect the M8008A Clock Generator Sample Clock Out port with
both the M8199A AWG Clock In ports using clock cable (red cable).
Figure 14
on page -38 shows how to connect M8008A clock generator
module to the two M8199A AWG modules:
Figure 14
Connecting M8008A clock generator to two M8199A AWG modules
M8100A Series Arbitrary Waveform Generators Getting Started Guide