Setup Procedure
Connecting to a Satellite Phone
Thermo Scientific
1. From the Admin menu, select Serial Mode/Speed > Reachback (see

Figure 22: Selecting Reachback in Admin menu

2. The Reachback Phone# option displays. From the Admin menu, select Reachback
Phone# (see
Figure

Figure 23: Reachback Phone # in Admin menu

3. Enter the Reachback phone number using the keypad. See your system
administrator for the correct number.
CAUTION Reachback phone number can contain numbers, *, #, and commas.
Do not enter alphanumeric characters "A-Z" or "a-z". If you do, the software
may freeze and require rebooting of the instrument.
1. Connect the RIIDEye Reachback cable, with its white label and female USB
connector, to the serial connection on RIIDEye (see
a. Match the red dot on the cable to the red dot on the round serial connector.
b. Push the cable connector straight into the RIIDEye connector – do not twist.
2. Connect the USB connector on the Reachback cable to the Iridium phone as
follows:
a. For 9555:
i. Connect the phone directly to the Reachback cable's USB connector.
b. For 9505a:
i. Connect the phone's interface cable (supplied with phone) to the
Reachback cable's USB connector.
12/22/2016
23).
RIIDEye DB-099E Operator Manual
Figure
Figure
21).
2
22).
22