Installing / removing and connecting analyzer and option modules
7.9 Removing OXYMAT 7, high-temperature version
7.9

Removing OXYMAT 7, high-temperature version

Overview
Figure 7-10
Procedure
1. Isolate the wall-mounted device from power.
2. Loosen the clamping ring connections at the gas inlets and outlets.
3. Loosen the six screws
78
Wall-mounted device
Screw for door (6 units)
Screw for housing floor (8 units)
Removing OXYMAT 7 high-temperature
Flat seal
Equipotential bonding cable
Analyzer module OXYMAT 7, high-temperature ver‐
sion
and open the door of the wall housing.
Operating Instructions, 07/2023, A5E31930403-AB