8.2 Electrical connections
1. Prepare the Ethernet cable:
Figure 8-10
– Isolate the Ethernet cable
– Insert the Ethernet cable through the cable gland. Make sure that the exposed shielding
– Screw the cable gland closed.
2. Fasten the prepared gland at the device:
Figure 8-11
– Fasten the Ethernet cable in the cable gland
Connecting the Ethernet cable
1. Isolate the free cable end.
2. Assemble the stripped cores in a FastConnect plug.
3. Insert the Ethernet plug into the service interface of the processing module. You must feel for
the position of the interface.
112
Cable gland
②④
Contact shielding
Ethernet cable
Shielding braid
Ethernet cable shielding
a length of approx. 1 cm.
braid rests on the shield contact
Plate with cable glands
Position cable gland Ethernet cable (underside of the device)
Ethernet cable: Position of the cable gland
at the desired location so that the braid
or
.
at the plate
Operating Instructions, 07/2023, A5E31930403-AB