Maintenance and servicing
11.10 Eliminating faults and errors
4. Loosen the three screws
5. Loosen the screw
bottom of the housing.
6. Remove the complete cable gland set.
7. Install the new complete cable gland set
Note
Fastening the cable glands
Fasten the cable glands (except conduit)
• M16 x 1.5 clamping range 5 to 10 mm made of plastic, counter nut and cap nut (clamp
• M20 x 1.5 clamping range 10 to 14 mm made of plastic, counternut and cap nut with a
• M16 x 1.5 clamping range 6 to 10 mm made of metal, counternut and cap nut with a
• M20 x 1.5 clamping range 10 to 14 mm made of metal, counternut and cap nut with a
Tightening the cable glands (conduit)
Metal adapter M32x1 on NPT1, tighten the locknut with a torque of 35 Nm. (Connecting
signal cables (Page 107))
8. Tighten the screw
9. Remount the shield plate. The torque at the screws
10.Close the door using six screws.
11.10
Eliminating faults and errors
Table 11-1
Error messages
Fault message
Description
Invalid device configu‐
No analyzer module was detected. At least one analyzer module is
ration
required to use the device.
Firmware versions in‐
The firmware of the basic device and the analyzer modules must be
compatible
compatible with one another.
Unknown module
An analyzer module has been detected which is not supported in the
type detected
firmware version.
154
②
③
on the inside of the right side panel and the eleven hex nuts at the
cables or plugs) with a torque of 3.0 Nm.
torque of 4.5 Nm.
torque of 6.5 Nm.
torque of 9.0 Nm.
③
again with a torque of 2.0 Nm.
and remove the shield plate
④
with eleven hex nuts with a torque of 0.8 Nm.
Operating Instructions, 07/2023, A5E31930403-AB
①
.
②
amounts to 0.8 Nm.
Measures
Install at least one analyzer
module.
Please contact the service de‐
partment.
Please contact the service de‐
partment.
Wall-mounted device