Performing installation
①
②
Figure 7-16
1. Inserting the terminal block:
– Insert the "Options" terminal block
– Tighten the 4 screws with a torque of 2.0 Nm.
2. Connect both terminal blocks with the existing connecting cable
3. Install the cable gland plate including both terminal blocks
with a torque of 0.8 Nm.
4. Tighten the screw for the cable gland plate again with a torque of 2.0 Nm.
Note
You use the same procedure for the conduit connections as for the cable gland plate.